Fertility and sterility · 2011
Essure transcervical tubal sterilization: a 5-year x-ray follow up.
Контролируемое клиническое исследованиеСтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Franchini Mario, Boeri Cecilia, Calzolari Stefano, Imperatore Alberto, Cianferoni Luciano, Litta Pietro, Giarrè Giovanna, Zerbetto Irene, Moncini Irene, Florio Pasquale
Аннотация
OBJECTIVE: To evaluate the x-ray appearance of Essure microinserts 5 years after their insertion.
DESIGN: Prospective controlled study.
SETTING: Tertiary referral centers for gynecologic care.
PATIENT(S): Forty-five consecutive women with successful hysteroscopic bilateral placement of the Essure devices and postprocedure satisfactory hysterosalpingography confirmation test.
INTERVENTION(S): Pelvic anteroposterior x-ray.
MAIN OUTCOME MEASURE(S): Stability and and symmetric appearance of Essure microinsert positions; measurement of the intrauterine distance between the two devices.
RESULT(S): After 5 years from their placement, no detachment nor fracture of devices was observed. x-Ray recognition of the device after 5 years showed findings similar to those recorded at 3 months' follow-through hysterosalpingography.
CONCLUSION(S): x-Ray evaluation of findings related to stability of position, symmetric appearance, and distance between the two Essure microinserts corroborates the irreversibility and the reliability of the fibrotic reaction that ensured tubal occlusion after devices placement.
DESIGN: Prospective controlled study.
SETTING: Tertiary referral centers for gynecologic care.
PATIENT(S): Forty-five consecutive women with successful hysteroscopic bilateral placement of the Essure devices and postprocedure satisfactory hysterosalpingography confirmation test.
INTERVENTION(S): Pelvic anteroposterior x-ray.
MAIN OUTCOME MEASURE(S): Stability and and symmetric appearance of Essure microinsert positions; measurement of the intrauterine distance between the two devices.
RESULT(S): After 5 years from their placement, no detachment nor fracture of devices was observed. x-Ray recognition of the device after 5 years showed findings similar to those recorded at 3 months' follow-through hysterosalpingography.
CONCLUSION(S): x-Ray evaluation of findings related to stability of position, symmetric appearance, and distance between the two Essure microinserts corroborates the irreversibility and the reliability of the fibrotic reaction that ensured tubal occlusion after devices placement.