MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1080/17576180.2026.2677733

Медицинская библиотека · Bioanalysis · 2026

Quantifying diclofenac - proven in theory, problematic in practice.

Hofmann Sarah, Schöning Eva, Huber Peter, Leger Felix, Wiese Heike

СтатьяОбзорПеревод в очереди

Аннотация

AIM: During clinical sample analysis unforeseen problems arose using a previously validated assay for the quantification of diclofenac in the concentration range from 0.0500 to 50.0 ng/mL. Here, we describe the troubleshooting process leading to the identification of the root cause and resulting changes in assay development strategy for robust supported liquid extraction (SLE)-based assays for clinical application.

METHODS

Individual steps of the sample preparation workflow and subsequent analysis with liquid chromatography and mass spectrometry (LCMS) were examined. Troubleshooting results were tested using stored validation samples and pooled clinical samples for confirmation.

RESULTS

The root cause for altered assay performance could be identified as the SLE plate batch effect attributable to variability in the sorbent material (diatomaceous earth). Reducing the sample loading volume applied to the SLE plates yielded purer extracts and consistent chromatographic performance. Quantitative recovery and signal consistency were restored after the modification.

CONCLUSION

Assay robustness was improved by underloading SLE plates. After adjusting the method and re-validation, clinical samples were successfully analyzed. The internal method development strategy was adjusted to avoid full capacity loading of SLE plate with natural sorbents.

Для цитирования: Hofmann Sarah, Schöning Eva, Huber Peter, Leger Felix, Wiese Heike. Quantifying diclofenac - proven in theory, problematic in practice.. Bioanalysis. 2026. DOI: 10.1080/17576180.2026.2677733.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.