MedElevator Войти в исследования
← Поиск публикаций

DOI 10.3785/j.issn.1008-9292.2008.02.009

Медицинская библиотека · Zhejiang da xue xue bao. Yi xue ban = Journal of Zhejiang University. Medical sciences · 2008

[Determination of dimethyl sulphate residual in granisetron hydrochloride by headspace gas chromatography].

Deng Gui-feng, Yao Tong-wei

Аннотация на английскомСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVE

To develop a headspace gas chromatography method for determining dimethyl sulphate residual in granisetron hydrochloride.

METHODS

An Angilent INNOWAX capillary column with nitrogen gas as carrier and FID as detector was applied in this study. Dimethyl sulphate was tested under a constant column temperature.

RESULT: Dimethyl sulphate had different retention time from other organic solvents such as alcohol,acetoacetate, isopropanol, dichlormethane and chloroform, which might exist in granisetron hydrochloride. The detection limit of dimethyl sulphate;s was 0.0016%.

CONCLUSION

The method can be used for the determination of dimethyl sulphate residual in granisetron hydrochloride.

Для цитирования: Deng Gui-feng, Yao Tong-wei. [Determination of dimethyl sulphate residual in granisetron hydrochloride by headspace gas chromatography].. Zhejiang da xue xue bao. Yi xue ban = Journal of Zhejiang University. Medical sciences. 2008. DOI: 10.3785/j.issn.1008-9292.2008.02.009.