Recanalization of thrombosed dialysis access with use of a rotating mini-pigtail catheter: follow-up study.

Schmitz-Rode T, Wildberger J E, Hübner D, Wein B, Schürmann K, Günther R W

Сравнительное исследованиеСтатьяПеревод в очереди

Аннотация

PURPOSE: To evaluate the feasibility, efficacy, and safety of mechanical thrombectomy of occluded dialysis access with use of a rotating mini-pigtail catheter.

MATERIALS AND

METHODS

Thrombus was fragmented by mechanical action of the rotating pigtail tip (5-mm diameter), while the guide wire exited a sidehole at the pigtail curvature and served as a fixed rotation axis. Twenty-six procedures were performed in 22 patients (12 men, 10 women; mean age, 55.5 years). Native fistulas were treated in 15 instances, polytetrafluoroethylene (PTFE) grafts were treated in 11 instances. Average occlusion time was 20 hours +/- 13 (range, 5-46 hours), average occlusion length was 25.6 cm +/- 10.1 (range, 6-45 cm). Thrombus fragmentation was followed by balloon angioplasty of underlying stenoses.

RESULTS

In all 26 procedures, the dialysis access was successfully declotted with subsequent dialysis using the access (clinical success rate, 100%). Handling of the mini-pigtail catheter was simple and rapid, regardless of whether a graft or a native fistula was treated. Average duration of the intervention was 118 minutes +/- 30. Mean primary patency was 165 days +/- 167. Primary patency rate was 82% at 30 days, 65% at 3 months, and 47% at 6 months. There was no evidence of complications due to the thrombus fragmentation procedure.

CONCLUSION

The results suggest that declotting of occluded dialysis grafts and fistulas with the mini-pigtail catheter is as effective and safe as other more established percutaneous therapies. It may serve as an easy-to-handle, low-budget alternative to current thrombectomy devices.

Для цитирования: Schmitz-Rode T, Wildberger J E, Hübner D, Wein B, Schürmann K, Günther R W. Recanalization of thrombosed dialysis access with use of a rotating mini-pigtail catheter: follow-up study.. Journal of vascular and interventional radiology : JVIR. 2000. DOI: 10.1016/s1051-0443(07)61630-3.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.