DOI 10.1016/s0003-4975(00)02636-9
Lessons learned from the first clinical implants of the DeBakey ventricular assist device axial pump: a single center report.
Аннотация
BACKGROUND
The bridge to transplantation with pulsatile mechanical assist devices became a standard procedure for patients deteriorating on the waiting list. Recently, continuous flow axial impeller pumps were introduced to clinical application offering new advantages.
METHODS
From November 1998 till September 2000, 6 male patients (mean age 53 plus or minus 11 years) with end-stage left heart failure were implanted with a DeBakey ventricular assist device (VAD) axial-flow pump for bridge to transplantation.
RESULTS
Three patients were successfully transplanted after 74, 115, and 117 days, respectively. Two other patients died after 25 and 133 days. One patient is still on the device after 108 days. Because of modification of the implantation technique after the first 2 patients, mean pump-flow within the first 3 weeks was increased from 4.3 +/- 0.6 L/min to 6.7 +/- 0.3 L/min. Patients were put on regular bicycle-ergometer training and improved their exercise capacities up to a mean maximum oxygen consumption of 20.2 mL/kg/min.
CONCLUSIONS
Initial implants of the DeBakey VAD demonstrated support properties comparable to pulsatile pumps but without significant restrictions for extended use.