DOI 10.1002/jcla.23216

Медицинская библиотека · Journal of clinical laboratory analysis · 2020

Identification of and solution for false D-dimer results.

Zhang Xian-Yan, Zhang Xue-Xuan, Xu Jia-Long, Huang Teng-Yi, Wu Ying, Yang Ye-Ru, Zhou Huan-Bin, Wu Ying-E

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Clinically, D-dimer (DD) levels are mainly used to exclude diseases such as deep venous thrombosis (DVT). In clinical testing, DD assays can be subjected to interference that may cause false results, which directly affect the clinical diagnosis. Our hypothesis was that the 95% confidence intervals (CIs) of the fibrin degradation product (FDP)/DD and fibrinogen (Fib)/DD ratios were used to identify these false results and corrected via multiple dilutions.

METHODS

In total, 16 776 samples were divided into three groups according to the DD levels detected by Sysmex CS5100 and CA7000: Group A, DD ≥ 2.0 μg/mL fibrinogen equivalent unit (FEU); group B, 0.5 < DD < 2.0 μg/mL FEU; and group C, DD ≤ 0.5 μg/mL FEU. The 95% CIs of the FDP/DD and Fib/DD ratios were calculated. Six abnormal DD results were found according to the 95% CIs. For verification, we performed multiple dilutions, compared the results with those of other instruments, and tested the addition of heterophilic blocking reagent (HBR).

RESULTS

The median and 95% CI of the FDP/DD ratio were 3.76 and 2.25-8.15 in group A, 5.63 and 2.86-10.58 in group B, 10.23 and 0.91-47.71 in groups C, respectively. For the Fib/DD ratio, the 95% CIs was 0.02-2.21 in group A, 0.68-8.15 in group B, and 3.82-55.27 in groups C. Six abnormal results were identified after multiple dilutions, by comparison with other detection systems, and after HBR addition.

CONCLUSIONS

The FDP/DD ratio is more reliable for identifying false results. If the FDP/DD ratio falls outside the 95% CI, it should be verified by different methods.

Для цитирования: Zhang Xian-Yan, Zhang Xue-Xuan, Xu Jia-Long, Huang Teng-Yi, Wu Ying, Yang Ye-Ru, Zhou Huan-Bin, Wu Ying-E. Identification of and solution for false D-dimer results.. Journal of clinical laboratory analysis. 2020. DOI: 10.1002/jcla.23216.