Comparison of Left Ventricular Pressure-Volume Loops With Veno-Arterial Extracorporeal Membrane Oxygenator Support: In Vitro Investigation.
СтатьяСравнительное исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Wu Eric L, Boesch Johannes, Nonaka Hideaki, Djelovic Christopher, Wickramarachchi Avishka, Benitez Jessica, Fraser John F
Аннотация
BACKGROUND: The hemodynamic consequence of increasing veno-arterial extracorporeal membrane oxygenator (V-A ECMO) flow on left ventricular (LV) loading is controversial, with in vivo studies reporting conflicting results on left ventricular distension. We investigated increased V-A ECMO flow on LV pressure-volume characteristics under varying cardiac and volume conditions using a repeatable in vitro mock circulation loop (MCL).
METHODS: A MCL was configured with V-A ECMO system to simulate four cardiogenic shock patients due to decompensated chronic heart failure patient: (1) LV failure with healthy right ventricle (RV) at normal volume (mean circulatory filling pressure, MCFP: 7 mmHg), (2) combined LV and RV failure at normal volume, (3) LV failure with healthy RV and volume overload (MCFP: 12 mmHg), and (4) combined LV and RV failure with volume overload. For each condition, ECMO flow was increased from 2 L/min to 4 L/min, and systemic vascular resistance (SVR) was adjusted to evaluate effects on LV pressure-volume loops.
RESULTS: In LV failure with preserved RV function with normal volume, increased ECMO flow at constant SVR led to LV distension (> 30 mL increase in end-diastolic volume). Reducing SVR mitigated this effect but resulted in hypotension (MAP < 65 mmHg). In combined LV and RV failure at normal volume, LV distension was less pronounced and manageable with SVR reduction while maintaining adequate MAP. Under volume overload, ECMO flow increases caused marked LV dilation and elevated central venous pressure, risking venous congestion.
CONCLUSIONS: We demonstrated that LV distension during V-A ECMO flow is influenced by ventricular function, volume status, and SVR. Judicious volume management and afterload control are essential, and patients with preserved RV function may require adjunct LV unloading strategies.
METHODS: A MCL was configured with V-A ECMO system to simulate four cardiogenic shock patients due to decompensated chronic heart failure patient: (1) LV failure with healthy right ventricle (RV) at normal volume (mean circulatory filling pressure, MCFP: 7 mmHg), (2) combined LV and RV failure at normal volume, (3) LV failure with healthy RV and volume overload (MCFP: 12 mmHg), and (4) combined LV and RV failure with volume overload. For each condition, ECMO flow was increased from 2 L/min to 4 L/min, and systemic vascular resistance (SVR) was adjusted to evaluate effects on LV pressure-volume loops.
RESULTS: In LV failure with preserved RV function with normal volume, increased ECMO flow at constant SVR led to LV distension (> 30 mL increase in end-diastolic volume). Reducing SVR mitigated this effect but resulted in hypotension (MAP < 65 mmHg). In combined LV and RV failure at normal volume, LV distension was less pronounced and manageable with SVR reduction while maintaining adequate MAP. Under volume overload, ECMO flow increases caused marked LV dilation and elevated central venous pressure, risking venous congestion.
CONCLUSIONS: We demonstrated that LV distension during V-A ECMO flow is influenced by ventricular function, volume status, and SVR. Judicious volume management and afterload control are essential, and patients with preserved RV function may require adjunct LV unloading strategies.