Медицинская библиотека · International orthopaedics · 1996
Reconstruction femoral nailing for nonunion of subtrochanteric fracture: a revision technique following dynamic condylar screw failure.
Клинический случайСтатьяПеревод в очереди
Аннотация
We report two cases of nonunion of subtrochanteric fractures associated with failure of fixation with a dynamic condylar screw (DCS) due to plate fracture. In both cases the nonunion was successfully treated by insertion of a reconstruction femoral nail. This implant allowed early weightbearing despite the patients' advanced age and poor bone quality. We did not perform direct bone grafting and the fractures healed without complication.
Для цитирования: Charnley G J, Ward A J. Reconstruction femoral nailing for nonunion of subtrochanteric fracture: a revision technique following dynamic condylar screw failure.. International orthopaedics. 1996. DOI: 10.1007/s002640050028.