MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/pace.12431

Медицинская библиотека · Pacing and clinical electrophysiology : PACE · 2014

Clinical, anatomical, and technical risk factors for postoperative pacemaker or defibrillator lead perforation with particular focus on myocardial thickness.

Schwerg Marius, Stockburger Martin, Schulze Christoph, Bondke Hansjürgen, Poller Wolfram C, Lembcke Alexander, Melzer Christoph

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Postoperative lead perforation is a life-threatening complication of cardiac pacing. Identification of precipitating factors for this serious complication may help to anticipate a specific risk profile and to minimize the incidence.

METHODS

We conducted a retrospective tertiary referral center analysis to clarify clinical, anatomical, and technical characteristics related to pacemaker (PM) and cardioverter/defibrillator lead perforation. We examined the baseline characteristics and the symptoms. In a subgroup, we investigated the myocardial thickness on contrast-enhanced cardiac computed tomography.

RESULTS

We enrolled 26 patients. Female gender appears to put patients at slightly increased risk for lead perforation. In a majority active fixation leads were used. Symptoms occurred in 72%. Pericardial effusion and tamponade were present in 38% and 19%, respectively. Sensing was compromised in 65%. A high pacing threshold or exit block occurred in 92%. Myocardial thickness did not differ between patients with or without perforation. In 96%, the perforation was treated by transvenous withdrawal.

CONCLUSION

Chest pain, phrenic stimulation, bad sensing, or exit block early after PM implantation must prompt radiological and echocardiographic evaluation. A missing pericardial effusion particularly late after implantation does not rule out a perforation. Especially active fixating leads have a higher risk of perforation. With cardiac surgery in standby transvenous withdrawal is a safe way to treat lead perforation.

Для цитирования: Schwerg Marius, Stockburger Martin, Schulze Christoph, Bondke Hansjürgen, Poller Wolfram C, Lembcke Alexander, Melzer Christoph. Clinical, anatomical, and technical risk factors for postoperative pacemaker or defibrillator lead perforation with particular focus on myocardial thickness.. Pacing and clinical electrophysiology : PACE. 2014. DOI: 10.1111/pace.12431.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.