Mechanical failure of the intramedullary hip screw in a subtrochanteric femoral fracture.

Rantanen J, Aro H

Клинический случайСтатьяПеревод в очереди

Аннотация

The Gamma nail, an implant specifically designed for intertrochanteric and subtrochanteric femoral fractures, has been criticized for its high risk of secondary shaft fractures. A modified design, the intramedullary hip screw has recently been introduced to correct this complication. We present a case of mechanical failure of this new implant that occurred in a pathological subtrochanteric fracture. The centering sleeve of the implant became loose and migrated while the head screw penetrated the acetabulum. Refixation was successful.

Для цитирования: Rantanen J, Aro H. Mechanical failure of the intramedullary hip screw in a subtrochanteric femoral fracture.. Journal of orthopaedic trauma. 1996. DOI: 10.1097/00005131-199607000-00010.