Geometry-Related Technical Complications After LVAD Implantation: Impact on Device Dysfunction and Clinical Outcomes.

Merzah Ali Saad, Deniz Ezin, Knigge Sara, Hanke Jasmin, Zubarevich Alina, Dia Mohamad, Sofia Chaikovska, Borys Todurov, Torabi Saeed, Schmack Bastian, Weymann Alexander, Darbaidze Zurab, Haunschild Josephina, Etz Christian D, Biewer Max, Ruhparwar Arjang, Schmitto Jan D, Dogan Günes

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Geometry of LVAD implantation is a determinant of pump performance and hemocompatibility. Despite technically successful operations, avoidable geometric errors may produce characteristic imaging patterns and clinically relevant device dysfunction.

OBJECTIVES

To determine the frequency, diagnostic pathway, and clinical impact of preventable geometry-related failure modes after HeartMate 3 implantation in a single-center referral cohort, specifically inflow cannula malposition, outflow graft kinking/excessive length, and aortic anastomotic stenosis.

METHODS

Retrospective, single-center analysis of 21 consecutive HeartMate 3 recipients implanted between 2016 and 2025 at external centers and subsequently evaluated in our department. All patients underwent CT/CTA when device dysfunction was suspected. Failure modes were predefined and adjudicated using standardized imaging criteria. CT/CTA was the primary modality used to confirm geometric failure modes following triggers such as low-flow alarms, hemolysis, and abnormal chest radiography.

RESULTS

Among 21 HeartMate 3 recipients, inflow cannula malposition was identified in 7 patients, outflow graft kinking/excessive length in 12, and aortic anastomotic stenosis in 2.

CONCLUSIONS

In this single-center referral cohort, geometric failure modes were identified in 15 of 21 patients (71.4%) after HeartMate 3 implantation and showed reproducible CT/CTA signatures. A standardized imaging-driven diagnostic pathway may improve recognition of correctable geometric problems in patients referred for suspected device dysfunction; given the referral-based design, these frequencies should not be interpreted as the true incidence in an unselected HeartMate 3 population.

Для цитирования: Merzah Ali Saad, Deniz Ezin, Knigge Sara, Hanke Jasmin, Zubarevich Alina, Dia Mohamad, Sofia Chaikovska, Borys Todurov, Torabi Saeed, Schmack Bastian, Weymann Alexander, Darbaidze Zurab, Haunschild Josephina, Etz Christian D, Biewer Max, Ruhparwar Arjang, Schmitto Jan D, Dogan Günes. Geometry-Related Technical Complications After LVAD Implantation: Impact on Device Dysfunction and Clinical Outcomes.. Artificial organs. 2026. DOI: 10.1111/aor.70204.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.