Impact of Laser Power and Firing Angle on Coagulation Efficiency in Laser Treatment for Twin-Twin Transfusion Syndrome: An ex vivo Placenta Study.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Akkermans Joost, van der Donk Loes, Peeters Suzanne H P, van Tuijl Sjoerd, Middeldorp Johanna M, Lopriore Enrico, Oepkes Dick
Аннотация
OBJECTIVE: To assess the impact of laser power and firing angle on coagulation efficiency for closing placental anastomoses in the treatment of twin-twin transfusion syndrome.
METHODS: We used an ex vivo blood-perfused human placenta model to compare time to complete coagulation using 30 vs. 50 W of neodymium-doped yttrium aluminum garnet laser power and using a firing angle of 90° vs. 45°. Placentas were perfused with pig blood at 5 mL/min. Differences were analyzed using independent-samples t test, Mann-Whitney U test, or χ2 test as appropriate.
RESULTS: Coagulation took less time and energy using 50 W (n = 53) compared to 30 W (n = 52), 11 vs. 22 s (p < 0.001), and 557 vs. 659 J (p = 0.007). Perpendicular coagulation (n = 53) took less time and energy compared to a 45° angle (n = 21), 11 vs. 17 s (p = 0.004), and 557 vs. 871 J (p = 0.004). Bleeding complicated 2 (3%) measurements in the 50-W group, 5 (10%) in the 30-W group, and 3 (14%) in the 45° group.
DISCUSSION: In a highly controlled model, a 50-W laser power setting was more energy efficient than 30 W in coagulating a placental vein. A more perpendicular laser firing angle resulted in more efficient coagulation. Furthermore, bleeding due to vessel wall disruption occurred more often with lower power and a more tangential approach.
METHODS: We used an ex vivo blood-perfused human placenta model to compare time to complete coagulation using 30 vs. 50 W of neodymium-doped yttrium aluminum garnet laser power and using a firing angle of 90° vs. 45°. Placentas were perfused with pig blood at 5 mL/min. Differences were analyzed using independent-samples t test, Mann-Whitney U test, or χ2 test as appropriate.
RESULTS: Coagulation took less time and energy using 50 W (n = 53) compared to 30 W (n = 52), 11 vs. 22 s (p < 0.001), and 557 vs. 659 J (p = 0.007). Perpendicular coagulation (n = 53) took less time and energy compared to a 45° angle (n = 21), 11 vs. 17 s (p = 0.004), and 557 vs. 871 J (p = 0.004). Bleeding complicated 2 (3%) measurements in the 50-W group, 5 (10%) in the 30-W group, and 3 (14%) in the 45° group.
DISCUSSION: In a highly controlled model, a 50-W laser power setting was more energy efficient than 30 W in coagulating a placental vein. A more perpendicular laser firing angle resulted in more efficient coagulation. Furthermore, bleeding due to vessel wall disruption occurred more often with lower power and a more tangential approach.