DOI 10.3113/fai.2010.1006

Медицинская библиотека · Foot & ankle international · 2010

Over-the-top knot placement technique enhances tensile stability of tendon repairs.

Türker Mehmet, Cetik Ozgür, Kılıçoğlu Onder, Cırpar Meriç, Dirikolu Hüsnü, Kalaycıoğlu Barş, Oztürk Latif

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Currently a major concern for the surgical treatment of Achilles tendon rupture repairs is the creation of stable enough fixation to allow early range of motion. It was documented that the weakest point in a suture loop is the knot. Thus, we hypothesized that moving the knot away from the repair junction (over-the-top Krackow technique) would increase the strength of the repair.

MATERIALS AND

METHODS

Transected bovine tendons were repaired by the traditional Krackow and over-the-top Krackow techniques using four suture materials (Fiberwire Nos. 5 and 2, Ethibond Nos. 5 and 2). Tendons were cyclically tested at incremental loads beginning from 50 N until 5-mm gap formation. Then all tendons were loaded to failure. The number of cycles to 5-mm gapping, ultimate failure loads and knot slip were compared using t-test and Mann-Whitney tests (with Tukey corrections for multiple comparisons).

RESULTS

Mean number of cycles to 5-mm gapping did not reveal significant differences (p = 0.113) between repair groups. Mean failure load of tendons repaired by over-the-top Krackow technique were significantly higher (p < 0.0001) for all four paired groups than tendons repaired by traditional Krackow technique. Ethibond No. 5, No. 2, and Fiberwire No. 2 suture repairs with over-the-top configuration did not reveal any knot slip.

CONCLUSION

Over-the-top Krackow technique increases the ultimate failure load of repaired tendons. But 5-mm gapping resistivity was not enhanced either by the technique or the suture material.

CLINICAL RELEVANCE: The knot itself is a stress-riser in the suture loop so we suggest that freeing it from tension by our modificiation may achieve more durable repairs.

Для цитирования: Türker Mehmet, Cetik Ozgür, Kılıçoğlu Onder, Cırpar Meriç, Dirikolu Hüsnü, Kalaycıoğlu Barş, Oztürk Latif. Over-the-top knot placement technique enhances tensile stability of tendon repairs.. Foot & ankle international. 2010. DOI: 10.3113/fai.2010.1006.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.