DOI 10.1016/j.joms.2020.02.025
Osteosynthesis of the Mandibular Condyle With Magnesium-Based Biodegradable Headless Compression Screws Show Good Clinical Results During a 1-Year Follow-Up Period.
Аннотация
PURPOSE: The use of titanium-based implants in mandibular condyle fractures can require implant removal because of screw penetration through the condylar surface. The use of biodegradable implants can avoid a second operation for implant removal and the associated possible complications. We investigated the clinical and radiologic outcomes of osteosynthesis of mandibular condyle fractures (MCFs) with biodegradable magnesium-based compression screws.
MATERIALS AND
METHODS
We performed a retrospective observational study of 6 patients who had been treated at our department. We recorded the changes in jaw movements over time, occlusion, and possible complications at defined intervals of 1, 3, 6, and 12 months postoperatively. We also compared the preoperative computed tomography (CT) scans with the postoperative cone-beam CT (CBCT) scans at 6 and 12 months postoperatively to evaluate mandibular condyle healing and screw degradation.
RESULTS
Of the 6 patients, 4 were men and 2 were women, with a mean age of 43.2 years (range, 30 to 66 years). All 6 patients had unilateral MCFs. All the patients showed well-restored function of the temporomandibular joint with significant improvement in mouth opening (46.17 ± 6.49 mm), right (10.67 ± 1.03 mm) and left (10.67 ± 1.97 mm) laterotrusion, and protrusion (10.17 ± 1.33 mm) distances to physiologic values. The CBCT scans showed the remodeling processes of the mandibular condyle and a few radiolucencies indicating the magnesium-based screws. Although penetration of 1 screw tip through the condylar surface had occurred, no implant removal was necessary owing to biodegradation of the implant.
CONCLUSIONS
The results of the present study have shown that biodegradable magnesium-based compression screws provide good clinical results and avoid implant removal.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.