MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/corr.0000000000002985

Медицинская библиотека · Clinical orthopaedics and related research · 2024

Femoral Component Design Is Associated With the Risk of Periprosthetic Femur Fracture After Cementless THA in Patients 65 Years or Older.

Kelly Mackenzie, Chen Antonia F, Ryan Sean P, Working Zachary M, De Ayushmita, Mullen Kyle, Porter Kimberly R, Kagan Ryland

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Periprosthetic femur fracture is a known complication after THA. The associated risk of cementless femoral component design for periprosthetic femur fracture in a registry population of patients older than 65 years has yet to be clearly identified.

QUESTIONS/PURPOSES: (1) Is femoral stem geometry associated with the risk of periprosthetic femur fracture after cementless THA? (2) Is the presence or absence of a collar on cementless femoral implant designs associated with the risk of periprosthetic femur fracture after THA?

METHODS

We analyzed American Joint Replacement Registry data from 2012 to March 2020. Unique to this registry is the high use of cementless femoral stems in patients 65 years and older. We identified 266,040 primary cementless THAs during the study period in patients with a diagnosis of osteoarthritis and surgeries linked to supplemental Centers for Medicare and Medicaid data where available. Patient demographics, procedure dates, and reoperation for periprosthetic femur fracture with revision or open reduction and internal fixation were recorded. The main analysis was performed comparing the Kheir and Chen classification: 42% (112,231 of 266,040) were single-wedge, 22% (57,758 of 266,040) were double-wedge, and 24% (62,983 of 266,040) were gradual taper/metadiaphyseal-filling cementless femoral components, which yielded a total of 232,972 primary cementless THAs. An additional analysis compared cementless stems with collars (20% [47,376 of 232,972]) with those with collarless designs (80% [185,596 of 232,972]). A Cox proportional hazard regression analysis with the competing risk of death was used to evaluate the association of design and fracture risk while adjusting for potential confounders.

RESULTS

After controlling for the potentially confounding variables of age, sex, geographic region, osteoporosis or osteopenia diagnosis, hospital volume, and the competing risk of death, we found that compared with gradual taper/metadiaphyseal-filling stems, single-wedge designs were associated with a greater risk of periprosthetic femur fracture (HR 2.9 [95% confidence interval (CI) 2.2 to 3.9]; p < 0. 001). Compared with gradual taper/metadiaphyseal-filling stems, double-wedge designs showed an increased risk of periprosthetic femur fracture (HR 3.0 [95% CI 2.2 to 4.0]; p < 0. 001). Collarless stems showed an increased risk of periprosthetic fracture compared with collared stems (HR 7.8 [95% CI 4.1 to 15]; p < 0. 001).

CONCLUSION

If cementless femoral fixation is used for THA in patients 65 years or older, surgeons should consider using gradual taper/metadiaphyseal-filling and collared stem designs because they are associated with a lower risk of periprosthetic femur fracture. Future investigations should compare gradual taper/metadiaphyseal-filling and collared cementless designs with cemented fixation in this population.

LEVEL OF EVIDENCE: Level III, therapeutic study.

Для цитирования: Kelly Mackenzie, Chen Antonia F, Ryan Sean P, Working Zachary M, De Ayushmita, Mullen Kyle, Porter Kimberly R, Kagan Ryland. Femoral Component Design Is Associated With the Risk of Periprosthetic Femur Fracture After Cementless THA in Patients 65 Years or Older.. Clinical orthopaedics and related research. 2024. DOI: 10.1097/corr.0000000000002985.