Biomechanical testing of different osteosynthesis systems for segmental resection of the mandible.

Schupp Wiebke, Arzdorf Michael, Linke Berend, Gutwald Ralf

СтатьяПеревод в очереди

Аннотация

PURPOSE: This investigation assessed the mechanical behavior of 3 different locking and nonlocking reconstruction systems-Unilock 2.4, Reconstruction 2.4, and Reconstruction 2.7-with regard to plate and screw fracture.

MATERIALS AND

METHODS

Five different plate/screw configurations (Unilock 2.4-locking screws, Unilock 2.4 -conventional screws, Reconstruction 2.4-conventional screws, Reconstruction 2.7-conventional screws, and Unilock 2.4-locking screws with a 1-mm gap; Synthes, Umkirch, Germany) were tested on synthetic mandibles. All mandibles were resected on the left side between the canine and third molar, reconstructed, and loaded cyclically between 30 and 300 N up to 250,000 cycles or until screw or plate failure occurred.

RESULTS

No screw fractures were observed. All plates fractured close to the distal fragment. The Unilock plates fixed with locking screws withstood significantly more cycles until failure than the Reconstruction plates 2.4 fixed with conventional MF-Cortex screws. No significant differences were found in the other groups. Only 2 of the 34 plates tested, both of the Reconstruction 2.7 system, reached the runout limit.

CONCLUSIONS

Unilock plates fixed with locking screws have a higher long-term stability than the Reconstruction 2.4 system. A 1-mm gap between the plate and mandible does not lead to early screw failure in the Unilock 2.4 system with locking screws. The Reconstruction 2.7 system seems superior if well contoured, because 2 of those plates reached the runout limit; however, this system is not as easy to handle as the 2.4 systems, and good contouring is difficult to achieve. Therefore, we consider the Unilock 2.4 system with locking screws the best choice.

Для цитирования: Schupp Wiebke, Arzdorf Michael, Linke Berend, Gutwald Ralf. Biomechanical testing of different osteosynthesis systems for segmental resection of the mandible.. Journal of oral and maxillofacial surgery : official journal of the American Association of Oral and Maxillofacial Surgeons. 2007. DOI: 10.1016/j.joms.2006.06.306.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.