Anesthesiology · 2009

Cervical spine motion: a fluoroscopic comparison of the AirTraq Laryngoscope versus the Macintosh laryngoscope.

Сравнительное исследованиеСтатьяРандомизированное исследованиеПоддержка исследования: вне правительства СШАНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Turkstra Timothy P, Pelz David M, Jones Philip M

Аннотация

BACKGROUND: The optimal technique to intubate the trachea in patients presenting with a potential or documented cervical spine (C-spine) injury remains unresolved. Using continuous fluoroscopic video assessment, C-spine motion during laryngoscopy with an AirTraq Laryngoscope (King Medical Systems, Newark, DE) was compared to that with intubation using a Macintosh blade.

METHODS: Twenty-four healthy surgical patients gave written consent to participate in a crossover randomized controlled trial; all patients were subjected to both Macintosh and AirTraq laryngoscopy with manual inline stabilization after induction of anesthesia. The C-spine motion was examined at four areas: the occiput-C1 junction, C1-C2 junction, C2-C5 motion segment, and C5-thoracic motion segment. The time required for laryngoscopy was also measured.

RESULTS: C-spine motion using the AirTraq was less than that during Macintosh laryngoscopy, averaging 66% less (P < 0.01) at three of the motion segments studied, occiput-C1, C2-C5, and C5-thoracic. There was no difference at the C1-C2 segment. There was no significant difference in the time to accomplish laryngoscopy between the two devices.

CONCLUSIONS: For patients in whom C-spine movement is undesirable, use of the AirTraq Laryngoscope may be useful to limit movement without an increase in the duration of intubation.