MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1002/ccd.27726

Медицинская библиотека · Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions · 2018

Fracture in drug-eluting stents increases focal intimal hyperplasia in the atherosclerosed rabbit iliac artery.

Conway Claire, Desany Gerard J, Bailey Lynn R, Keating John H, Baker Brian L, Edelman Elazer R

СтатьяПеревод в очереди

Аннотация

OBJECTIVES

Drug-eluting stent (DES) strut fracture (SF) is associated with higher incidence of In-stent restenosis (ISR)-return of blockage in a diseased artery post stenting-than seen with bare metal stents (BMS). We hypothesize that concomitance of drug and SF leads to greater neointimal response.

BACKGROUND

Controlled release of therapeutic agents, such as sirolimus and its analogs, or paclitaxel from has reduced tissue based DES failure modes compared to BMS. ISR is dramatically reduced and yet the implications of mechanical device failure is magnified.

METHODS

Bilateral Xience Everolimus-eluting stents (EES) were implanted in 20 New Zealand White rabbits on normal (n = 7) or high fat (HF)/high cholesterol (HC) (n = 13) diets. Implanted stents were intact or mechanically fractured. Everolimus concentration was as packaged or pre-eluted. After 21 days, stented vessels were explanted, resin embedded, MicroCT scanned, and analyzed histomorphometrically.

RESULTS

Fractured EES were associated with significant (P < 0.05) increases in arterial stenosis and neointimal formation and lower lumen-to-artery area ratios compared to intact EES. Hyperlipidemic animals receiving pre-eluted EES revealed no significant difference between intact and fracture groups.

CONCLUSIONS

SF increases intimal hyperplasia, post EES implant, and worse with more advanced disease. Pre-eluted groups, reflective of BMS, did not show significant differences, suggesting a synergistic effect of everolimus and mechanical injury, potentially explaining the lack of SF reports for BMS. Here, we report that ISR has a higher incidence with SF in EES, the clinical implication is that patients with SF after DES implantation merit careful follow-up.

Для цитирования: Conway Claire, Desany Gerard J, Bailey Lynn R, Keating John H, Baker Brian L, Edelman Elazer R. Fracture in drug-eluting stents increases focal intimal hyperplasia in the atherosclerosed rabbit iliac artery.. Catheterization and cardiovascular interventions : official journal of the Society for Cardiac Angiography & Interventions. 2018. DOI: 10.1002/ccd.27726.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.