DOI 10.2147/mder.s626574

Медицинская библиотека · Medical devices (Auckland, N.Z.) · 2026

Agreement and Accuracy of a Commercial Pulse Oximeter Compared with Arterial Oxygen Saturation During Controlled Hypoxemia in Healthy Volunteers.

Lin Jinqi, Pan Shiyan, Ma Lan, Ye Jilun, Liu Chunsheng, Wang Fan, Zhang Xu

СтатьяПеревод в очереди

Аннотация

PURPOSE: Pulse oxygen saturation (SpO2) is widely used for continuous non-invasive oxygenation monitoring, but its accuracy can be affected by sensor design, probe placement, peripheral perfusion, skin pigmentation, motion artifacts, and device algorithms. This study evaluated the accuracy and agreement of a commercial pulse oximetry system (model M201 SpO2 module, Shenzhen Witleaf Medical Electronics Co. Ltd.) with a matched fingertip sensor during non-motion controlled hypoxemia in healthy volunteers, using arterial oxygen saturation (SaO2) measured by arterial blood gas/co-oximetry analysis as the reference standard.

MATERIAL AND

METHODS

Eighteen healthy volunteers underwent two phases of controlled hypoxemia. Each phase comprised three target SpO2 plateau ranges (90-100%, 80-89%, and 70-79%). Phase 1 was the desaturation phase, and Phase 2 was a re-oxygenation and repetition phase. After each target plateau stabilized, SpO2 values from the investigational system were recorded, and paired arterial blood samples were collected for SaO2 measurement. Measurement error was defined as SpO2-SaO2. Bias, mean absolute error (MAE), accuracy root mean square (ARMS), 95% limits of agreement (LOA), and proportions of absolute errors exceeding predefined thresholds were calculated.

RESULTS

A total of 432 paired SpO2-SaO2 observations were analyzed. SpO2 showed strong correlation with SaO2 (Pearson r = 0.990; Spearman r = 0.986; both p < 0.001). Overall bias was -0.061%, with 95% LOA from -2.65% to 2.53%. MAE and ARMS were 1.05% and 1.32%, respectively. Absolute errors exceeding ±2%, ±3%, and ±4% occurred in 11.57%, 2.55%, and 0% of observations, respectively. ARMS values were 1.03%, 1.19%, and 1.66% across the 90-100%, 80-89%, and 70-79% platform phases, respectively, with increased error dispersion and mild SaO2 underestimation in the 70-79% phase.

CONCLUSION

The commercial pulse oximetry system showed good overall agreement with reference SaO2 during controlled hypoxemia, with an ARMS of 1.32% (well below the 4% threshold). Error dispersion increased in the 70-79% saturation range, with mild underestimation of SaO2 by SpO2. These findings support the clinical performance of this device within the tested saturation range.

Для цитирования: Lin Jinqi, Pan Shiyan, Ma Lan, Ye Jilun, Liu Chunsheng, Wang Fan, Zhang Xu. Agreement and Accuracy of a Commercial Pulse Oximeter Compared with Arterial Oxygen Saturation During Controlled Hypoxemia in Healthy Volunteers.. Medical devices (Auckland, N.Z.). 2026. DOI: 10.2147/mder.s626574.