Foetal Radiation Dose During Prophylactic Occlusion Balloon Placement for Morbidly Adherent Placenta.
СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Semeraro Vittorio, Susac Ana, Morasca Andrea, D'Antonio Francesco, Belli Anna-Maria
Аннотация
OBJECTIVE: To assess foetal absorbed radiation dose (FAD) during prophylactic occlusion balloon catheter (POBC) placement before surgery for morbidly adherent placenta (MAP).
METHODS: Retrospective cohort study of women with MAP undergoing POBC before surgery. Women were divided into two groups. Group 1 treated between 2008 and 2011, Group 2 between 2012 and 2014 when fluoroscopy pulse rate reduced from 7.5 to 2 pulses per second. Median and interquartile range (IQR) of dose area product (DAP) and FAD were compared using the Mann-U-Whitney test. FAD was calculated using the PCXMC Monte-Carlo calculation. All procedures were performed using the same equipment (Siemens Axiom Artis dTA) by the same experienced interventional radiology team.
RESULTS: Thirty-four women underwent POBC. 16 procedures were performed in Group 1 and 18 in Group 2. Median DAP overall was 1025.10 µGym(2) (IQR 532.1-1775.3): Group 1, 1713.25 µGym(2) (IQR 1164.5-2274.5) and Group 2, 660.70 µGym(2) (IQR 440.9-1020.9) (p = 0.027). Median FAD overall was 4.65 mGy (IQR 2.7-8.2): Group 1, 6.25 mGy (IQR 4.4-10.6) and Group 2, 3.05 mGy (IQR 2.0-5.4) (p = 0.031).
CONCLUSIONS: FAD during POBC insertion is low overall. The only change between the two groups was in pulsed fluoroscopy rate which resulted in a significant reduction in FAD. This reinforces the importance of implementation of dose reduction techniques.
METHODS: Retrospective cohort study of women with MAP undergoing POBC before surgery. Women were divided into two groups. Group 1 treated between 2008 and 2011, Group 2 between 2012 and 2014 when fluoroscopy pulse rate reduced from 7.5 to 2 pulses per second. Median and interquartile range (IQR) of dose area product (DAP) and FAD were compared using the Mann-U-Whitney test. FAD was calculated using the PCXMC Monte-Carlo calculation. All procedures were performed using the same equipment (Siemens Axiom Artis dTA) by the same experienced interventional radiology team.
RESULTS: Thirty-four women underwent POBC. 16 procedures were performed in Group 1 and 18 in Group 2. Median DAP overall was 1025.10 µGym(2) (IQR 532.1-1775.3): Group 1, 1713.25 µGym(2) (IQR 1164.5-2274.5) and Group 2, 660.70 µGym(2) (IQR 440.9-1020.9) (p = 0.027). Median FAD overall was 4.65 mGy (IQR 2.7-8.2): Group 1, 6.25 mGy (IQR 4.4-10.6) and Group 2, 3.05 mGy (IQR 2.0-5.4) (p = 0.031).
CONCLUSIONS: FAD during POBC insertion is low overall. The only change between the two groups was in pulsed fluoroscopy rate which resulted in a significant reduction in FAD. This reinforces the importance of implementation of dose reduction techniques.