MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1136/fetalneonatal-2013-305665

Медицинская библиотека · Archives of disease in childhood. Fetal and neonatal edition · 2014

Bi-level CPAP does not improve gas exchange when compared with conventional CPAP for the treatment of neonates recovering from respiratory distress syndrome.

Lampland Andrea L, Plumm Brenda, Worwa Cathy, Meyers Patricia, Mammel Mark C

Сравнительное исследованиеСтатьяРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

AIM: We hypothesised that short-term application of bi-level nasal continuous positive airway pressure CPAP (SiPAP) compared with conventional nasal CPAP (nCPAP) at the same mean airway pressure in infants with persistent oxygen need recovering from respiratory distress syndrome would improve CO2 removal with no change in oxygen requirement.

DESIGN

Non-blinded, randomised, observational four-period crossover study.

SETTING/POPULATION: Level III NICU; low-birthweight infants requiring CPAP and oxygen while recovering from respiratory distress syndrome.

METHODS

Infants requiring nasal CPAP for >24 h prior to study enrolment, and fraction of inspired oxygen requirement (FiO2) of 0.25-0.5, were randomised to either nCPAP or SiPAP. A crossover design with four 1 h treatment periods was used such that each infant received both treatments twice. Oxygen saturations (SaO2), transcutaneous CO2 (tcCO2) and vital signs were monitored continuously. Polysomnographic recordings were analysed for apnoea, bradycardia and oxygen desaturation.

RESULTS

Twenty low-birthweight infants receiving 0.3±0.04% supplemental oxygen on CPAP of 6 cm H2O were studied at an average of 33 days of age (±23 days, SD). There were no differences in tcCO2 or other physiological parameters except mean blood pressure, which was lower during nCPAP (52.3±8.3 vs 54.4±9.1 mm Hg; ±SD; p<0.01). No differences in short or prolonged apnoea, bradycardia or significant desaturation events were observed.

CONCLUSIONS

At similar mean airway pressures, SiPAP does not improve CO2 removal, oxygenation or other studied physiological parameters with the exception of mean blood pressure, which was not clinically significant.

TRIAL REGISTRATION NUMBER: NCT01053455.

Для цитирования: Lampland Andrea L, Plumm Brenda, Worwa Cathy, Meyers Patricia, Mammel Mark C. Bi-level CPAP does not improve gas exchange when compared with conventional CPAP for the treatment of neonates recovering from respiratory distress syndrome.. Archives of disease in childhood. Fetal and neonatal edition. 2014. DOI: 10.1136/fetalneonatal-2013-305665.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.