Safety of LigaSure in recurrent laryngeal nerve dissection-porcine model using continuous monitoring.
Аннотация
OBJECTIVES/HYPOTHESIS: This study investigated recurrent laryngeal nerve (RLN) real-time electromyography (EMG) data to define optimal safety parameters of the LigaSure Small Jaw (LSJ) instrument during thyroidectomy.
STUDY
DESIGN
Prospective animal model.
METHODS
Dynamic EMG tracings were recorded from 32 RLNs (16 piglets) during various applications of LSJ around using continuous electrophysiologic monitoring. At varying distances from the RLN, the LSJ was activated (activation study). The LSJ was also applied to the RLN at timed intervals after activation and after a cooling maneuver through placement on the sternocleidomastoid muscle (cooling study).
RESULTS
In the activation study, there was no adverse EMG event at 2 to 5 mm distance (16 RLNs, 96 tests). In the cooling study, there was no adverse EMG event after 2-second cooling time (16 RLNs, 96 tests) or after the LSJ cooling maneuver on the surrounding muscle before reaching the RLNs (8 RLNs, 24 tests).
CONCLUSION
Based on EMG functional assessment, the safe distance for LSJ activation was 2 mm. Further LSJ-RLN contact was safe if the LSJ was cooled for more than 2 seconds or cooled by touch muscle maneuver. The LSJ should be used with these distance and time parameters in mind to avoid RLN injury.
LEVEL OF EVIDENCE: N/A. Laryngoscope, 127:1724-1729, 2017.