DOI 10.1016/j.clinbiomech.2024.106181

Медицинская библиотека · Clinical biomechanics (Bristol, Avon) · 2024

A functional brace for treating Colles fractures; volunteer trial of a new design.

Wardlaw Douglas, Murdoch Ann H

Клиническое исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Presently used Colles' fracture treatments have similar outcomes with significant complications. Previous studies of a dynamic functional fracture brace, achieving similar or better results had no significant complications. A novel brace design is described to achieve optimal patient outcomes.

METHODS

Patient forearm measurements combined with data from a previously tested brace provided parameters for a brace formed using computer assisted design. Fracture swelling was simulated using a 40 ml (ml) plastic water filled bag placed over the dorsum of the lower radius. The interface pressure between the brace, skin and bag were measured on twelve human volunteers, four with forearm measurements to fit each of small, medium and large brace sizes. Measurements were taken at intervals reducing volume of fluid in the bag by 10 ml until empty, first wearing a brace, and then with a Colles plaster of Paris cast.

FINDINGS: The brace produced interface pressures over the lower radius in the range of 62.5-90 (mm Hg) and when the bag was completely empty the pressure range was 43-83 mmHg. The initial interface pressures in the Colles casts were in the range of 15-18 mm of Mercury (mm Hg) and when 10 mls were removed from the plastic bag the interface pressure dropped immediately to almost zero.

INTERPRETATION: The brace measurements show pressures higher than a Colles cast which compensate for reduction of simulated swelling. It is safe and is potentially the optimal design to achieve significantly improved patient outcomes.

Для цитирования: Wardlaw Douglas, Murdoch Ann H. A functional brace for treating Colles fractures; volunteer trial of a new design.. Clinical biomechanics (Bristol, Avon). 2024. DOI: 10.1016/j.clinbiomech.2024.106181.