Long-Term Safety with Sling Mesh Implants for Stress Incontinence.

Chughtai Bilal, Mao Jialin, Matheny Michael E, Mauer Elizabeth, Banerjee Samprit, Sedrakyan Art

СтатьяПеревод в очереди

Аннотация

PURPOSE: We examined long-term risks and predictors of mesh erosion and reoperation following mid urethral sling procedure for stress urinary incontinence.

MATERIALS AND

METHODS

Women aged 18 years or older who received a mid urethral sling for stress urinary incontinence between 2008 and 2016 in outpatient surgical settings in New York State were included in our study. Those who underwent concomitant mesh pelvic organ prolapse repair were excluded. Primary outcomes were post-implantation time to erosion and reoperations. Kaplan-Meier analysis and Cox proportional hazard models were used to assess the risks of erosion diagnosis and reoperation.

RESULTS

Our cohort included 36,195 women with a mean±SD age of 53.7±12.4 years. Estimated risks of erosions and reoperations at 7 years after sling procedures were 3.7% and 6.7%, respectively. Older age (≥65 vs <65: HR 0.83, 95% CI 0.70-0.99) and high volume facilities (high vs low: HR 0.79, 95% CI 0.68-0.92) were associated with a lower risk of erosion. History of hysterectomy was associated with a higher risk of erosion (HR 1.62, 95% CI 1.36-1.92). Predictors of reoperation included concurrent abdominal or native tissue transvaginal prolapse repair, previous hysterectomy and depression.

CONCLUSIONS

One in 27 women had sling erosions and 1 in 15 had invasive reoperations at 7 years after sling procedures. The highest erosion cases were observed among younger White women treated at low volume facilities. Continued and vigilant surveillance of mesh in stress urinary incontinence repairs, the nature and burden of stress urinary incontinence recurrence, different types of re-treatment, patient reported outcomes and information about treating surgeons are crucial.

Для цитирования: Chughtai Bilal, Mao Jialin, Matheny Michael E, Mauer Elizabeth, Banerjee Samprit, Sedrakyan Art. Long-Term Safety with Sling Mesh Implants for Stress Incontinence.. The Journal of urology. 2020. DOI: 10.1097/ju.0000000000001312.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.