DOI 10.1016/j.jacep.2024.09.025

Медицинская библиотека · JACC. Clinical electrophysiology · 2024

Safety and Feasibility of Pulsed-Field Ablation in Patients With Mechanical Prosthetic Valves.

Scacciavillani Roberto, Della Rocca Domenico G, Nair Devi, Magnocavallo Michele, Głowniak Andrzej, Mohanty Sanghamitra, Sorgente Antonio, Mené Roberto, Almorad Alexandre, Vetta Giampaolo, Pannone Luigi, Wojewoda Katarzyna, Marcon Lorenzo, Stroker Erwin, Bala Gezim, Konopka Aleksander, Sieira Juan, Bianchi Stefano, Gabrah Kirollos, Rossi Pietro, Sarkozy Andrea, Boveda Serge, Natale Andrea, de Asmundis Carlo, Chierchia Gian-Battista

СтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Pulsed-field ablation (PFA) is a novel ablation energy source, which leads to selective and irreversible electroporation of cardiomyocytes with no collateral damage to adjacent tissues. The presence of mechanical prosthetic valves may pose some issues related to catheter manipulation and risk of electromagnetic interference.

OBJECTIVES

The purpose of this study was to assess the feasibility of PFA in patients with mechanical prosthetic valves.

METHODS

The authors enrolled 30 consecutive patients (age 64.7 ± 11.4 years; 43% men) with mitral and/or aortic valve replacement with mechanical substitutes undergoing endocardial AF ablation via a PFA technology at 6 institutions.

RESULTS

Time between valve surgery and ablation was 6.6 ± 6.5 years; 14 (46.7%) patients had an aortic mechanical valve, 10 (33.3%) a mitral, and 6 (20.0%) both. The following PFA technologies were used: FARAWAVE in 25 patients, Sphere-9 in 2, PulseSelect in 2, and the CENTAURI System in 1. No periprocedural and 1-month postablation major and minor complications were observed. Postprocedural transthoracic echocardiography did not identify any prosthetic valve dysfunction. One patient showed electromagnetic interference of the PFA catheter adjacent to the mitral prosthesis without clinical implications. No energy delivery to the mitral edge of the isthmus was possible despite multiple attempts to reposition the catheter. Mean procedural time was 77.5 ± 29.9 minutes, whereas mean left atrial dwell and fluoroscopy times were 43.1 ± 21.4 and 14.0 ± 6.4 minutes, respectively.

CONCLUSIONS

PFA in patients with mechanical valves appears safe, but careful catheter handling is essential to avoid electromagnetic interference hindering PFA delivery.

Для цитирования: Scacciavillani Roberto, Della Rocca Domenico G, Nair Devi, Magnocavallo Michele, Głowniak Andrzej, Mohanty Sanghamitra, Sorgente Antonio, Mené Roberto, Almorad Alexandre, Vetta Giampaolo, Pannone Luigi, Wojewoda Katarzyna, Marcon Lorenzo, Stroker Erwin, Bala Gezim, Konopka Aleksander, Sieira Juan, Bianchi Stefano, Gabrah Kirollos, Rossi Pietro, Sarkozy Andrea, Boveda Serge, Natale Andrea, de Asmundis Carlo, Chierchia Gian-Battista. Safety and Feasibility of Pulsed-Field Ablation in Patients With Mechanical Prosthetic Valves.. JACC. Clinical electrophysiology. 2024. DOI: 10.1016/j.jacep.2024.09.025.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.