Langer Jessica M, Cohen Raphael M, Berns Jeffrey S, Chittams Jesse, Cooper Emily T, Trerotola Scott O
Аннотация
PURPOSE: Over-the-wire exchange of tunneled dialysis catheters is the standard of care per K/DOQI guidelines for treating catheter-related bacteremia. However, Gram-positive bacteremia, specifically with staphylococcus species, may compromise over-the-wire exchange due to certain biological properties. This study addressed the effectiveness of over-the-wire exchange of staphylococcus-infected tunneled dialysis catheters compared with non-staphylococcus-infected tunneled dialysis catheters.
METHODS: Patients who received over-the-wire exchange of their tunneled dialysis catheter due to documented or suspected bacteremia were identified from a QA database. Study patients (n = 61) had positive cultures for Staphylococcus aureus, Staphylococcus epidermidis, or coagulase-negative staphylococcus not otherwise specified. Control patients (n = 35) received over-the-wire exchange of their tunneled dialysis catheter due to infection with any organism besides staphylococcus. Overall catheter survival and catheter survival among staphylococcal species were assessed.
RESULTS: There was no difference in tunneled dialysis catheter survival between study and control groups (P = 0.46). Median survival time was 96 days for study catheters and 51 days for controls; survival curves were closely superimposed. There also was no difference among the three staphylococcal groups in terms of catheter survival (P = 0.31). The median time until catheter removal was 143 days for SE, 67 days for CNS, and 88 days for SA-infected catheters.
CONCLUSIONS: There is no significant difference in tunneled dialysis catheter survival between over-the-wire exchange of staphylococcus-infected tunneled dialysis catheters and those infected with other organisms.
METHODS: Patients who received over-the-wire exchange of their tunneled dialysis catheter due to documented or suspected bacteremia were identified from a QA database. Study patients (n = 61) had positive cultures for Staphylococcus aureus, Staphylococcus epidermidis, or coagulase-negative staphylococcus not otherwise specified. Control patients (n = 35) received over-the-wire exchange of their tunneled dialysis catheter due to infection with any organism besides staphylococcus. Overall catheter survival and catheter survival among staphylococcal species were assessed.
RESULTS: There was no difference in tunneled dialysis catheter survival between study and control groups (P = 0.46). Median survival time was 96 days for study catheters and 51 days for controls; survival curves were closely superimposed. There also was no difference among the three staphylococcal groups in terms of catheter survival (P = 0.31). The median time until catheter removal was 143 days for SE, 67 days for CNS, and 88 days for SA-infected catheters.
CONCLUSIONS: There is no significant difference in tunneled dialysis catheter survival between over-the-wire exchange of staphylococcus-infected tunneled dialysis catheters and those infected with other organisms.