MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1186/1749-8090-7-36

Медицинская библиотека · Journal of cardiothoracic surgery · 2012

Right ventricular rupture and tamponade caused by malposition of the Avalon cannula for venovenous extracorporeal membrane oxygenation.

Hirose Hitoshi, Yamane Kentaro, Marhefka Gregary, Cavarocchi Nicholas

Клинический случайСтатьяПеревод в очереди

Аннотация

Placement of the Avalon Elite bicaval dual lumen cannula for venovenous extracorporeal membrane oxygenation (VV-ECMO) via the internal jugular vein requires precise positioning of the cannula tip in the inferior vena cava with echocardiography or fluoroscopy guidance. Correct guidewire placement is clearly the key first step in assuring proper advancement of the cannula. We report a case of unexpected wire migration into the right ventricle at the time of final cannula advancement, resulting in right ventricular rupture and tamponade. Transesophageal echocardiography is an important monitoring modality for appropriate placement of the VV-ECMO guidewire and Avalon cannula, and in particular, for early identification of potential complications.

Для цитирования: Hirose Hitoshi, Yamane Kentaro, Marhefka Gregary, Cavarocchi Nicholas. Right ventricular rupture and tamponade caused by malposition of the Avalon cannula for venovenous extracorporeal membrane oxygenation.. Journal of cardiothoracic surgery. 2012. DOI: 10.1186/1749-8090-7-36.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.