Subacromial Balloon Spacer Implantation for Massive Irreparable Rotator Cuff Tears Is Associated With Restoration of the Acromiohumeral Interval and Glenohumeral Center of Pressure: A Systematic Review and Meta-Analysis of Controlled Laboratory Studies.

Kunze Kyle N, Moran Jay, Taylor Samuel A, Fu Michael C, Rodeo Scott A, Warren Russell F, Dines David M, Gulotta Lawrence V, Dines Joshua S

МетаанализСистематический обзорСтатьяПеревод в очереди

Аннотация

BACKGROUND

Biodegradable subacromial balloon spacers (SBSs) have become increasingly used for the treatment of massive irreparable rotator cuff tears given their theorized clinical benefits; however, the relationship between biomechanical functions of the balloon spacer and clinical benefits remains unclear.

PURPOSE: To perform a systematic review and meta-analysis of controlled laboratory studies investigating the use of SBSs for massive irreparable rotator cuff tears.

STUDY

DESIGN

Systematic review and meta-analysis; Level of evidence, 4.

METHODS

PubMed, OVID/Medline, and Cochrane databases were queried in July 2022 for biomechanical data pertaining to SBS implantation in cadaveric models of irreparable rotator cuff tears. Random-effects meta-analysis of continuous outcomes using the DerSimonian-Laird method was performed to estimate pooled-treatment effect sizes between the irreparable rotator cuff tear state and the state in which an SBS was implanted. Data reported variably or in formats not amenable to analysis were presented descriptively.

RESULTS

Five studies involving 44 cadaveric specimens were included. At 0° of shoulder abduction, SBS implantation resulted in a mean inferior humeral head translation of 4.80 mm (95% CI, 3.20-6.40; P < .001) relative to the irreparable rotator cuff tear state. This decreased to 4.39 mm and 4.35 mm at 30° and 60° of abduction, respectively. At 0° of abduction, implantation of an SBS was associated with a 5.01-mm (95% CI, 3.56-6.46, P < .001) anterior translation of the glenohumeral center of contact pressure relative to the irreparable tear state. This translation changed to 5.11 mm and 5.49 mm at 30° and 60° of abduction. In 2 studies, SBS implantation restored the glenohumeral contact pressure to that of the intact state and significantly reduced subacromial pressure distribution over a rotator cuff repair state. In 1 study, a high balloon fill volume (40 mL) resulted in a significant 10.3 ± 1.4-mm more anterior humeral head position relative to the intact cuff state.

CONCLUSION

SBS implantation in cadaveric models of irreparable rotator cuff tears results in significant improvements in humeral head position at 0°, 30°, and 60° of shoulder abduction. Balloon spacers may also improve glenohumeral and subacromial contact pressures, although insufficient evidence currently exists to corroborate these findings. High balloon fill volumes (40 mL) may confer supraphysiologic anteroinferior translation of the humeral head.

Для цитирования: Kunze Kyle N, Moran Jay, Taylor Samuel A, Fu Michael C, Rodeo Scott A, Warren Russell F, Dines David M, Gulotta Lawrence V, Dines Joshua S. Subacromial Balloon Spacer Implantation for Massive Irreparable Rotator Cuff Tears Is Associated With Restoration of the Acromiohumeral Interval and Glenohumeral Center of Pressure: A Systematic Review and Meta-Analysis of Controlled Laboratory Studies.. The American journal of sports medicine. 2023. DOI: 10.1177/03635465221150652.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.