DOI 10.1016/j.jacep.2025.11.016

Медицинская библиотека · JACC. Clinical electrophysiology · 2025

Visualization of PFA During PVI With the Second-Generation Pentaspline Catheter: NAVIGATE-PF Phase 1 Results.

García-Bolao Ignacio, Reddy Vivek Y, Su Wilber W, Koruth Jacob S, Fitzpatrick Noel, Neuzil Petr, Rincón Ramón Albarrán, Canepa Silvia, Kang Kaylen, Oesterlein Tobias, Koop Brendan E, Metzdorff Camille, Okeson Brynn, Gutbrod Sarah R, Szeplaki Gabor

СтатьяМногоцентровое исследованиеКлиническое исследование I фазыПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

BACKGROUND

Pulsed field ablation (PFA) for atrial fibrillation ablation provides a unique challenge for acute lesion evaluation due to reversible myocardial injury. Real-time guidance during ablation would provide reference for lesion location and extent. The second-generation pentaspline PFA catheter improves mapping integration with real-time visualization of catheter shape and previews the estimated ablative electric field.

OBJECTIVES

This study sought to evaluate the estimated shape and position of the acute tags relative to low-voltage borders from high-density, postablation maps.

METHODS

A multicenter, first-in-human study, NAVIGATE-PF (Feasibility Study on the FARAVIEW Technology), was conducted in 30 atrial fibrillation patients. Tags were placed following each application based on the shape of the estimated electric field. Post ablation, a high-quality, high-density voltage map was created with a high-density mapping catheter. Tags were overlaid on the high-density map and contours were drawn at the border of low voltage (≤0.5mV) and the outer border where at least 2 overlapping tags were placed.

RESULTS

All 30 patients were successfully treated with the second-generation PFA catheter. For the 15 patients included in the acute tags analysis, the region of acute electrical isolation correlated with the estimated ablative electric field. Post-procedural processing of the distance between the tag and low-voltage border was -0.58 mm (Q1-Q3: -2.9 to 2.17 mm) where a negative number indicates the tag is smaller than the low-voltage border.

CONCLUSIONS

The second-generation pentaspline PFA catheter, with dynamic shape visualization and preview of anticipated electric field, resulted in alignment with postablation voltage mapping. (Feasibility Study on the FARAVIEW Technology [NAVIGATE-PF]; NCT06175234).

Для цитирования: García-Bolao Ignacio, Reddy Vivek Y, Su Wilber W, Koruth Jacob S, Fitzpatrick Noel, Neuzil Petr, Rincón Ramón Albarrán, Canepa Silvia, Kang Kaylen, Oesterlein Tobias, Koop Brendan E, Metzdorff Camille, Okeson Brynn, Gutbrod Sarah R, Szeplaki Gabor. Visualization of PFA During PVI With the Second-Generation Pentaspline Catheter: NAVIGATE-PF Phase 1 Results.. JACC. Clinical electrophysiology. 2025. DOI: 10.1016/j.jacep.2025.11.016.