Influence of Mesh Design on Hemodynamic Performance and Embolic Filtration in Distal Embolic Protection Filters.

Sánchez-Nevárez Manuel Ignacio, Serra-Lluch Juan, Plana Emma, Quintero-Igeño Pedro, Medina Pilar, Miralles Manuel

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Distal net-type embolic protection filters (EPDn) are widely used during endovascular interventions to reduce embolic complications. However, the impact of mesh design on filtration performance and flow dynamics remains poorly defined.

METHODS

A controlled in vitro experimental model simulating arterial flow under physiological pulsatile conditions was developed. Four EPDn were evaluated. A standardized bolus of 140 μm microspheres was injected. Filtration performance was assessed using capture efficacy (CE), overall efficacy (OE), and particle leakage fraction (PLF), while hemodynamic impact was quantified by peripheral resistance units (PRU). Morphometric analysis and computational modeling were used to characterize mesh design.

RESULTS

Baseline resistance differed significantly among devices, with the highest PRU observed in EPDn1 (5.42 vs. 0.86-4.50; P < 0.001). After particle capture, EPDn4 showed the greatest increase in resistance (11.07 vs. 0.96-8.33; P < 0.001). CE was highest for EPDn4 (97.84%), followed by EPDn3 (78.67%), EPDn1 (63.39%), and EPDn2 (33.24%) (P < 0.001). PLF was lower in EPDn3 and EPDn4 (0.21 vs. 0.45-0.58; P < 0.01). OE was highest for EPDn4 (77.28%), followed by EPDn3 (62.75%), EPDn1 (35.27%), and EPDn2 (14.17%) (P ≤ 0.01). Structural and computational analyses identified design-related differences that may explain these findings.

CONCLUSION

Mesh design significantly influences the performance of EPDn. Devices with higher filtration efficiency may increase flow resistance, whereas others preserve hemodynamics at the expense of embolic protection. These trade-offs may have important implications for device selection in endovascular interventions.

Для цитирования: Sánchez-Nevárez Manuel Ignacio, Serra-Lluch Juan, Plana Emma, Quintero-Igeño Pedro, Medina Pilar, Miralles Manuel. Influence of Mesh Design on Hemodynamic Performance and Embolic Filtration in Distal Embolic Protection Filters.. Annals of vascular surgery. 2026. DOI: 10.1016/j.avsg.2026.07.028.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.