Corneal endothelial morphology and function after torsional and longitudinal ultrasound mode phacoemulsification.
Аннотация
PURPOSE: To study the endothelial cell morphology and corneal thickness changes after phacoemulsification by using the OZil torsional and longitudinal ultrasound techniques (Infiniti Vision System, Alcon Laboratories).
SETTING
Department of Ophthalmology, Clinical Center, University of Debrecen, Debrecen, Hungary.
METHODS
52 patients with cataract were randomly assigned to longitudinal ultrasound and torsional mode group. All surgeries were performed through a 2.2 mm clear corneal incision, the method employed being divide and conquer. The endothelial morphometry such as cell density (ECD), mean cell area, coefficient of variation of cell area, and central corneal thickness were examined with specular microscopy (EM-1000, Tomey) preoperatively and 4, 8 weeks postoperatively.
RESULTS
ECD values decreased significantly in both surgical groups (P < .001, repeated- mesures ANOVA), the postoperative endothelial cell loss was higher in the longitudinal ultrasound mode group (3.5% and 6.5%, at 4 and 8 weeks after surgery) than in the torsional group (3.3% and 5.5%, at 4 and 8 weeks after surgery), the difference not being significant between the two groups (P = .164 and P = .479, at 4 and 8 weeks after surgery, Mann-Whitney test). There was no statistically significant difference in any of the assessed parameters between the two surgical groups (P > .05). No significant correlation was found between the endothelial cell loss and the nucleus density.
CONCLUSIONS
Both phacoemulsification techniques were safe and effective. The torsional handpiece performs oscillatory movements and delivers less energy into the eye than the longitudinal ultrasound technique, therefore providing more favorable energy and thermal safety profile.