MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1177/08853282221125309

Медицинская библиотека · Journal of biomaterials applications · 2022

In vitro and in vivo assessment of a novel ultra-flexible ventriculoamniotic shunt for treating fetal hydrocephalus.

Emery Stephen P, Greene Stephanie, Elsisy Moataz, Chung Kaitlin, Ye Sang-Ho, Kim Seungil, Wagner William R, Hazen Nika, Chun Youngjae

СтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

Fetal aqueductal stenosis (AS) is one of the most common causes of congenital hydrocephalus, which increases intracranial pressure due to partial or complete obstruction of cerebrospinal fluid (CSF) flow within the ventricular system. Approximately 2-4 infants per 10,000 births develop AS, which leads to progressive hydrocephalus, which enlarges the head often necessitating delivery by cesarean section. Most babies born with AS are severely neurologically impaired and experience a lifetime of disability. Therefore, a new device technology for venticuloamniotic shunting is urgently needed and has been studied to ameliorate or prevent fetal hydrocephalus development, which can provide a significant impact on patients and their family's quality of life and on the decrease of the healthcare dollars spent for the treatment. This study has successfully validated the design of shunt devices and demonstrated the mechanical performance and valve functions. A functional prototype shunt has been fabricated and subsequently used in multiple in vitro tests to demonstrate the performance of this newly developed ventriculoamniotic shunt. The shunt contains a main silicone-nitinol composite tube, a superelastic 90° angled dual dumbbell anchor, and an ePTFE valve encased by a stainless-steel cage. The anchor will change its diameter from 1.15 mm (collapsed state) to 2.75 mm (deployed state) showing up to 1.4-fold diameter change in human body temperature. Flow rates in shunts were quantified to demonstrate the valve function in low flow rates mimicking the fetal hydrocephalus condition showing "no backflow" for the valved shunt while there is up to 15 mL/h flow through the shunt with pressure difference of 20 Pa. In vivo ovine study results show the initial successful device delivery and flow drainage with sheep model.

Для цитирования: Emery Stephen P, Greene Stephanie, Elsisy Moataz, Chung Kaitlin, Ye Sang-Ho, Kim Seungil, Wagner William R, Hazen Nika, Chun Youngjae. In vitro and in vivo assessment of a novel ultra-flexible ventriculoamniotic shunt for treating fetal hydrocephalus.. Journal of biomaterials applications. 2022. DOI: 10.1177/08853282221125309.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.