Cyclic joint loading can affect the initial stability of meniscal fixation implants.

Dürselen Lutz, Schneider Jan, Galler Mario, Claes Lutz E, Bauer Gerhard

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

OBJECTIVE

To identify the effect of cyclic joint loads on the initial fixation strength of different meniscal repair techniques.

DESIGN

Three different meniscal refixation implants and one suturing technique were tested for initial stability in porcine knee joints after cyclic joint loading.

BACKGROUND

Existing studies tested the initial pull out strength of meniscal fixation implants. It remains unknown how such implants perform under cyclic joint loading.

METHODS

The Meniscal Fastener, the Meniscal Arrow, the ClearFix Screw, and a horizontal suture (PDS 1) were implanted in 64 porcine knee joints after setting a tear in the posterior horn of the medial meniscus. In one group of each device the menisci were excised directly after implantation and subsequently subjected to a pull out test. The knee joints of the other groups were exposed to 5000 axial load cycles between 10 and 200 N prior to the pull out test.

RESULTS

The horizontal suture showed the highest pull out force (103 N, SD 19 N). Arrow (52 N, SD 18 N), Meniscal Fastener (29 N, SD 3 N), and Meniscal Screw (22 N, SD 8 N) failed at significantly lower loads. Cyclic loading led to a decreased initial pull out force only for the horizontal suture (82 N, SD 26 N) and the Meniscal Fastener (23 N, SD 5N).

CONCLUSIONS

Cyclic joint loading can lead to reduced fixation strength especially of meniscal refixation implants and sutures with relatively low stiffness. However, it remains unclear which fixation strength is necessary to provide conditions for meniscal healing.

RELEVANCE: The current study adds information to the understanding of how meniscal fixation systems perform mechanically especially after cyclic joint loads.

Для цитирования: Dürselen Lutz, Schneider Jan, Galler Mario, Claes Lutz E, Bauer Gerhard. Cyclic joint loading can affect the initial stability of meniscal fixation implants.. Clinical biomechanics (Bristol, Avon). 2003. DOI: 10.1016/s0268-0033(02)00139-0.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.