DOI 10.1007/s00392-022-02052-9

Медицинская библиотека · Clinical research in cardiology : official journal of the German Cardiac Society · 2022

Bioprosthetic valve dysfunction and failure after TAVI in bicuspid aortic valve stenosis during one-year follow-up according to VARC-3.

Veulemans Verena, Nuyens Philippe, Goh Shouheng, Maier Oliver, Binnebößel Stephan, Heermann Jacqueline, Jung Christian, Westenfeld Ralf, Kelm Malte, de Backer Ole, Zeus Tobias

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Transcatheter aortic valve implantation (TAVI) in bicuspid aortic valve (BAV) stenosis has become more frequent in the last years. This may pose challenges for long-time valve durability. Therefore, we aimed to evaluate the prevalence of bioprosthetic valve dysfunction (BVD) with the newest-generation devices in BAV stenosis up to one-year follow-up (FU).

METHODS

The primary endpoint was defined as the prevalence of BVD during the first procedural year according to Valve Academic Research Consortium (VARC)-3 criteria. Secondary endpoints were defined as failure in device success and clinical endpoints according to VARC-3.

RESULTS

A total of 107 patients were included. Of these, 34 subjects (31.8%) met the criteria for BVD during a mean FU of 263 ± 180 days, of which 20.2% were already documented after thirty days. Device success after one year was lower in the + BVD cohort (57.6% vs. 98.7%, p < 0.0001*). The rates of structural valve deterioration were 6.5%, non-structural valve deterioration (NSVD) 17.8%, subclinical leaflet thickening 10.3%, and endocarditis 0.9%. NSVD was foremost triggered by patient prosthesis mismatch in balloon-expandable valves. Hemodynamic valve deterioration stage 1 and 2 was confirmed in 16.8% of + BVD patients, while stage 1 and 3 bioprosthetic valve failure occurred in 1.9%. There was no impact of BVD on mortality.

CONCLUSION

There is critical evidence of early BVD after TAVI in BAV during one-year FU in one-third of patients, also lowering device success. The most frequently observed bioprosthetic valve dysfunction was NSVD due to patient prosthesis mismatch following TAVI with a balloon-expandable valve.

Для цитирования: Veulemans Verena, Nuyens Philippe, Goh Shouheng, Maier Oliver, Binnebößel Stephan, Heermann Jacqueline, Jung Christian, Westenfeld Ralf, Kelm Malte, de Backer Ole, Zeus Tobias. Bioprosthetic valve dysfunction and failure after TAVI in bicuspid aortic valve stenosis during one-year follow-up according to VARC-3.. Clinical research in cardiology : official journal of the German Cardiac Society. 2022. DOI: 10.1007/s00392-022-02052-9.