Journal of clinical microbiology · 2017
Multicenter Clinical Evaluation of BacT/Alert Virtuo Blood Culture System.
Оценочное исследованиеСтатьяМногоцентровое исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Jacobs Michael R, Mazzulli Tony, Hazen Kevin C, Good Caryn E, Abdelhamed Ayman M, Lo Pauline, Shum Bianche, Roman Katharine P, Robinson Danielle C
Аннотация
BacT/Alert Virtuo is an advanced, automated blood culture system incorporating improved automation and an enhanced detection algorithm to shorten time to detection. A multicenter study of the investigational Virtuo system (bioMérieux, Inc., Durham, NC) compared to BacT/Alert 3D (BTA3D) for detection of bacteremia/fungemia in four bottle types, SA and FA Plus (aerobic) and SN and FN Plus (anaerobic), was performed in a clinical setting with patient samples in a matched system design clinical trial. Blood was added to paired aerobic or anaerobic bottles, with the volume in each bottle in each pair required to be ≤10 ml and with the volumes required to be within 30% of each other. Of 5,709 bottle sets (52.5% aerobic pairs and 47.5% anaerobic pairs), 430 (7.5%) were positive for bacterial or fungal growth, with 342 (6.0%) clinically significant and 83 (1.5%) contaminated. A total of 3,539 sets (62.0%) were volume compliant, with 203 sets (5.7%) clinically significant. The positivity rates for volume-compliant bottle pairs determined by the two systems were comparable, with 68.7% of clinically significant isolates detected by both instruments, 15.7% by Virtuo only, and 15.7% by BTA3D only. Virtuo detected microbial growth nearly 2 h sooner overall than BTA3D (mean, 15.9 h versus 17.7 h). Shorter time to detection by Virtuo was related to organism group, with the time to detection being significantly shorter for enteric Gram-negative bacilli and enterococci (means, 3.6 h and 2.3 h shorter, respectively). This large clinical study demonstrated that the Virtuo blood culture system produced results comparable to those seen with the long-established BTA3D system, with significantly shorter time to detection.