MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jse.2017.09.037

Медицинская библиотека · Journal of shoulder and elbow surgery · 2017

The effect of a radial neck notch on press-fit stem stability: a biomechanical study on 7 cadavers.

Shukla Dave R, Sahu Dipit C, Fitzsimmons James S, An Kai-Nan, O'Driscoll Shawn W

СтатьяПеревод в очереди

Аннотация

BACKGROUND

Minimal micromotion is necessary for osteointegration of cementless radial head prostheses. When radial head fractures extend longitudinally, where the neck cut for prosthetic replacement should be made is uncertain. We hypothesized that complete resection of the notched portion of a radial neck confers no advantage in initial stability compared with not resecting the defect and inserting the implant into a notched radial neck.

MATERIALS AND

METHODS

The radii of 7 cadavers underwent radial head resection and implantation with a 25-mm-long press-fit radial head stem. Before implantation, a 5-mm-long notch that was less than 1-mm wide was made in the radial neck. After the stem-bone micromotion was recorded, the proximal 5 mm of radial neck, incorporating the entire notch, was cut away, the stem was inserted 5 mm further, and the resulting micromotion was recorded.

RESULTS

The mean micromotion measured in the presence of a cortical notch was 51 ± 6 µm. After the neck was circumferentially cut and the stem was advanced, the micromotion (46 ± 9 µm) was not statistically significantly different.

DISCUSSION: Initial stability of an adequately sized cementless stem in the presence of a 5-mm-long cortical notch was well within the threshold needed for bone ingrowth (<100 µm). In addition, there was no reduction of micromotion after the notch-containing portion of the radial neck was resected and the stem was advanced. Making a neck cut distal to a 5-mm-long, 1-mm-wide cortical notch does not confer added stability. Thus, surgeons can preserve bone stock and avoid an aggressive neck cut.

Для цитирования: Shukla Dave R, Sahu Dipit C, Fitzsimmons James S, An Kai-Nan, O'Driscoll Shawn W. The effect of a radial neck notch on press-fit stem stability: a biomechanical study on 7 cadavers.. Journal of shoulder and elbow surgery. 2017. DOI: 10.1016/j.jse.2017.09.037.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.