BACKGROUND

A patent foramen ovale (PFO) has been associated with embolic strokes and transient ischemic attacks (TIAs). Catheter closure of PFO is effective in preventing recurrent events. Residual shunts and procedure or device related complications can occur, including atrial fibrillation and thrombus formation. This study examines the initial experience with a new PFO closure device, the IrisFIT PFO-Occluder (Lifetech Scientific, Shenzhen, China).

METHODS

95 patients with indications for PFO closure underwent percutaneous closure with the IrisFIT PFO-Occluder. The primary endpoint was the rate of accurate device placement with no/small residual shunt at 3 or 6 months follow-up. All patients underwent transoesophageal echocardiography (TEE) after 1 to 6 months. In case of a residual shunt, an additional TEE was performed after 12 months. Clinical follow-up was performed up to a mean of 33.1 ± 3.6 months.

RESULTS

The device was successfully implanted in 95 (100%) patients with no relevant procedural complications. At final TEE follow-up (7.6 ± 3.9 months) the effective closure rate was 96.8% with 1 moderate and 2 large residual shunts. There were 8 cases of new onset atrial fibrillation and 2 TIAs. There were no cases of device embolization or erosion.

CONCLUSION

The IrisFIT occluder is a new PFO closure device with several advantages compared to other devices. In this small study cohort, technical success rate, closure rate and adverse event rate were comparable to other devices. The rate of new onset atrial fibrillation was higher in comparison to other studies and warrants further investigation.

Для цитирования: Sievert Kolja, Yu Jiangtao, Bertog Stefan, Hornung Marius, von Bardeleben Ralph Stephan, Gafoor Sameer, Reinartz Markus, Matic Predrag, Hofmann Ilona, Grunwald Iris, Schnelle Nalan, Sievert Horst. Post-Market Clinical Follow-Up With the Patent Foramen Ovale Closure Device IrisFIT (Lifetech) in Patients With Stroke, Transient Ischemic Attack, or Other Thromboembolic Events.. Cardiovascular revascularization medicine : including molecular interventions. 2020. DOI: 10.1016/j.carrev.2020.09.031.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.