DOI 10.1016/j.diagmicrobio.2011.06.001
Comparison of time to positivity of the VersaTREK® REDOX 80-mL and the REDOX EZ draw 40-mL blood culture bottles for common bacterial bloodstream pathogens.
Аннотация
The VersaTREK(®) microbial detection system offers 2 media formulations, an aerobic and an anaerobic bottle available in a 40-mL direct draw format and an 80-mL format. The 40-mL EZ Draw(®) bottle can be inoculated with a maximum volume of 5 mL, while the REDOX 80-mL bottle accommodates a 10-mL volume. The effect of volume of blood inoculum on time to positivity (TTP) has not been clearly established with these bottle types. This study utilized simulated blood cultures seeded with clinically relevant microorganisms in human blood to evaluate the impact of inoculum volume and organism load on TTP for the 2 bottle types. For 13/15 organisms, the EZ Draw bottle flagged positive earlier than the REDOX 80-mL bottles. The lower volume of blood inoculum did not negatively impact TTP using the EZ Draw blood culture bottles as compared to REDOX 80-mL bottles.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.