BACKGROUND

Intra-aortic balloon pumps (IABP) are commonly used as a bridge to heart transplant (Htx). Although IABPs were designed for transfemoral arterial placement, axillary IABPs (axIABP) allow for continued ambulation and rehabilitation. However, increased rates of device complications with axillary compared to femoral insertion suggest anatomic variations that predispose patients to device failure.

METHODS

We conducted a single-center retrospective study of patients who received axIABP as a bridge to HTx between February 2018 and July 2022. Pre-axIABP computed tomography (CT) images were reviewed for all patients, and anatomic features of the axillary and central arteries were measured and correlated with mechanical axIABP complications.

RESULTS

A total of 48 patients underwent successful axIABP placement with available pre-insertion CT images, of whom 35 (72.9%) were male. Out of 48 patients, 26 (54.2%) experienced 34 complications, including axIABP rupture (n = 10, 29%), infection (n = 8, 23%), and thrombosis (n = 6, 18%). CTs showed an average axillary artery diameter and depth of 6.79 and 55.6 mm, respectively. The average subclavian artery takeoff angle from the aortic arch was 95.4°.

CONCLUSION

Of the 10 vascular anatomic features measured, there was no association between any single measurement and axIABP complications. However, there may be other structural features not measured by routine CT, technical issues with insertion, and device factors not accounted for in this study that drive axIABP complications. Nonetheless, our findings suggest that normal variations in the measured vascular anatomy should not preclude axIABP use.

Для цитирования: Valdes Carlos Alberto, Drew Kristin, Sharaf Omar M, Ahmed Mustafa M, Parker Alex M, Vilaro Juan R, Aranda Juan M, Massoomi Michael R, Anderson R David, Jimenez Yomary, Al-Ani Mohammad A Z. Anatomic Suitability for Axillary Intra-Aortic Balloon Pump Circulatory Support.. Artificial organs. 2025. DOI: 10.1111/aor.14994.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.