Gauhar Vineet, Traxer Olivier, Kandarthanda Nanjappa Madappa, Somani Bhaskar K, Castellani Daniele, Sabnis Ravindra B, Ragoori Deepak
Аннотация
BACKGROUND: The direct in-scope suction (DISS) involves a two-way adaptor mounted on a scope to aspirate and irrigate the pelvicalyceal system during flexible ureteroscopy. While integrated suction single-use scopes manage small dust particles effectively, they cannot remove particles >250 µm and are limited by single use.
OBJECTIVES: To evaluate perioperative outcomes of the GLITZ system, a lightweight suction accessory for flexible ureteroscopes, facilitating laser lithotripsy with integrated aspiration.
DESIGN: Prospective, multicenter study.
METHODS: The GLITZ system, a 100 g trigger mounted on the ureteroscope handle, integrates with an irrigation and aspiration apparatus featuring a flow-regulating sensor. A finger-trigger activates suction, stopping irrigation until released, enhancing visibility and safety by automatically stopping irrigation if blockages occur. The study involved 29 patients (November 2023-April 2024). Procedures were performed using a disposable 7.5 Fr ureteroscope, ureteral access sheath, and thulium fiber or 100 W holmium:YAG laser. Surgeons evaluated ease of use and performance. Stone-free status was evaluated at a 30-day CT scan and classified as follows: Grade A: zero RF; Grade B: single RF not up 2 mm; Grade C: single RF 2.1-4 mm; and Grade D: single/multiple RFs > 4 mm. Data are reported as median and (interquartile range).
RESULTS: A total of 31.1% of patients were female. The median age was 54 years. Pain was the most common complaint, and 41.8% had multiple stones. GLITZ system showed satisfactory performance in 93.1% of cases, with complete dust aspiration achieved in 62.1%. Device dislodgement occurred in 24.1% of cases. Median operative time was 40 min, achieving postoperative CT at 30 days confirmed Grade A stone-free status in 75.9%. Fever was the most common complication (10.3%), with no sepsis case.
CONCLUSION: The GLITZ system in DISS shows effectiveness with a 96.6% stone-free rate (Grade A + B) at 30 days, indicating strong potential for flexible ureteroscopy, though additional studies are needed to optimize flow rates and usability.
OBJECTIVES: To evaluate perioperative outcomes of the GLITZ system, a lightweight suction accessory for flexible ureteroscopes, facilitating laser lithotripsy with integrated aspiration.
DESIGN: Prospective, multicenter study.
METHODS: The GLITZ system, a 100 g trigger mounted on the ureteroscope handle, integrates with an irrigation and aspiration apparatus featuring a flow-regulating sensor. A finger-trigger activates suction, stopping irrigation until released, enhancing visibility and safety by automatically stopping irrigation if blockages occur. The study involved 29 patients (November 2023-April 2024). Procedures were performed using a disposable 7.5 Fr ureteroscope, ureteral access sheath, and thulium fiber or 100 W holmium:YAG laser. Surgeons evaluated ease of use and performance. Stone-free status was evaluated at a 30-day CT scan and classified as follows: Grade A: zero RF; Grade B: single RF not up 2 mm; Grade C: single RF 2.1-4 mm; and Grade D: single/multiple RFs > 4 mm. Data are reported as median and (interquartile range).
RESULTS: A total of 31.1% of patients were female. The median age was 54 years. Pain was the most common complaint, and 41.8% had multiple stones. GLITZ system showed satisfactory performance in 93.1% of cases, with complete dust aspiration achieved in 62.1%. Device dislodgement occurred in 24.1% of cases. Median operative time was 40 min, achieving postoperative CT at 30 days confirmed Grade A stone-free status in 75.9%. Fever was the most common complication (10.3%), with no sepsis case.
CONCLUSION: The GLITZ system in DISS shows effectiveness with a 96.6% stone-free rate (Grade A + B) at 30 days, indicating strong potential for flexible ureteroscopy, though additional studies are needed to optimize flow rates and usability.