Comparative validation of speckle contrast optical spectroscopy against diffuse correlation spectroscopy for monitoring human cerebral blood flow.

СтатьяНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Cheng Tom Y, Robinson Mitchell B, Renna Marco, Wu Kuan-Cheng, Starkweather Zachary, Kierul Olivia S, Kim Byungchan Kenny, Howard Alexander C, Boas David A, Carp Stefan A, Cheng Xiaojun, Franceschini Maria Angela

Аннотация

SIGNIFICANCE: Within diffuse optics, speckle contrast optical spectroscopy (SCOS) has emerged as a promising alternative to the state-of-the-art technique of diffuse correlation spectroscopy (DCS) for continuous, noninvasive bedside monitoring of cerebral blood flow (CBF). Although the two methods theoretically yield equivalent relative indices of CBF (rCBFi), in practice, SCOS-based measurements require experimental calibration to obtain unbiased rCBFi values. To date, there are limited validation studies comparing SCOS and DCS in human subjects, particularly at long source-detector separations (SDS) relevant to adult brain monitoring.

AIM: We aim to compare rCBFi from SCOS and DCS during concurrent, colocalized measurements on tissue-mimicking phantoms and humans at a large SDS of 3 cm.

APPROACH: We conducted concurrent SCOS and DCS measurements on temperature-ramped, two-layer, and flow phantoms, respectively. We also conducted concurrent CBF measurements on 10 healthy volunteers undergoing various physiological challenges (breath-holding, hyperventilation, pressure modulation, and squatting) designed to elicit measurable changes in blood flow. SCOS and DCS operated from a shared pulsed laser source, which enabled pulsation-resolved human CBF measurements at 3 cm SDS (DCS necessitated the use of cardiac-gated averaging).

RESULTS: We observed strong agreement ( r = 0.93 , slope = 0.98) between SCOS- and DCS-derived rCBFi at 3 cm SDS and an order-of-magnitude improvement in SCOS noise performance relative to DCS during the in vivo measurements. Phantom experiments showed a small depth-sensitivity disadvantage for SCOS at the same SDS; however, this was far outweighed by its superior noise performance, yielding an order-of-magnitude improvement in SCOS contrast-to-noise ratio.

CONCLUSIONS: The results demonstrate the equivalence of properly calibrated SCOS and DCS rCBFi measurements at long SDS in adults, establishing SCOS as a viable alternative to DCS for monitoring CBF. Further validation across larger cohorts and clinical populations is warranted.