Temperature Sensor-Mediated Intraoperative Tachycardia in a Leadless Pacemaker During Cardiac Surgery.

Gokul Kaushik, Fernando Rohesh, Marchant Bryan, Royster Roger, Lata Adrian, Bradford Natalie, Bhave Prashant, Whalen Patrick, Bodziock George

Клинический случайСтатьяПеревод в очереди

Аннотация

BACKGROUND

Leadless pacemaker use has grown with some now including temperature-based sensors rather than motion-based sensors. We present a case of unexpected rewarming-induced rapid pacing during cardiopulmonary bypass in a patient with the Aveir leadless pacemaker.

CASE SUMMARY: We present a case of a patient with a leadless pacemaker undergoing cardiac surgery and suffering from rapid ventricular pacing when weaning from bypass treated by applying a magnet to stop rapid pacing with device interrogation conducted intraoperatively. The rate response feature was disabled with the patient weaned from bypass.

DISCUSSION: Traditional pacemakers use rate response features whereas the Aveir leadless pacemaker uses a temperature-based response feature. For this reason, the device is susceptible to inappropriate pacing when patients undergo rewarming during cardiac surgery resulting in possible hemodynamic compromise.

TAKE-HOME MESSAGE: Patients undergoing cardiac surgery with leadless pacemakers in place should have a careful preoperative evaluation with the rate response feature turned off before surgery if indicated.

Для цитирования: Gokul Kaushik, Fernando Rohesh, Marchant Bryan, Royster Roger, Lata Adrian, Bradford Natalie, Bhave Prashant, Whalen Patrick, Bodziock George. Temperature Sensor-Mediated Intraoperative Tachycardia in a Leadless Pacemaker During Cardiac Surgery.. JACC. Case reports. 2026. DOI: 10.1016/j.jaccas.2026.110001.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.