A comparison of non-invasive versus invasive measures of intracranial pressure in hypoxic ischaemic brain injury after cardiac arrest.

Cardim Danilo, Griesdale Donald E, Ainslie Philip N, Robba Chiara, Calviello Leanne, Czosnyka Marek, Smielewski Peter, Sekhon Mypinder S

Сравнительное исследованиеСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

AIM: Increased intracranial pressure (ICP) in hypoxic ischaemic brain injury (HIBI) can cause secondary ischaemic brain injury and culminate in brain death. Invasive ICP monitoring is limited by associated risks in HIBI patients. We sought to evaluate the agreement between invasive ICP measurements and non-invasive estimators of ICP (nICP) in HIBI patients.

METHODS

Eligible consecutive adult (age>18) cardiac arrest patients with HIBI were included as part of a single centre prospective interventional study. Invasive ICP monitoring and nICP measurements were undertaken using: a) transcranial Doppler ultrasonography (TCD), b) optic nerve sheet diameter ultrasound (ONSD) and c) jugular venous bulb pressure (JVP). Multiple measurements applied in linear mixed-effects models were considered to obtain the correlation coefficient between ICP and nICP as well as their predictive abilities to detect intracranial hypertension (ICP≥20mm Hg).

RESULTS

Eleven patients were included (median age of 47 [range 20-71], 8 males and 3 females). There was a linear relationship between ICP and nICP with ONSD (R=0.53 [p<0.0001]), JVP (R=0.38 [p<0.001]) and TCD (R=0.30 [p<0.01]). The ability to predict intracranial hypertension was highest for ONSD and TCD (area under the receiver operating curve (AUC)=0.96 [95% CI: 0.90-1.00] and AUC=0.91 [95% CI: 0.83-1.00], respectively). JVP presented the weakest prediction ability (AUC=0.75 [95% CI: 0.56-0.94]).

CONCLUSIONS

ONSD and TCD methods demonstrated agreement with invasively-monitored ICP, suggesting their potential roles in the detection of intracranial hypertension in HIBI after cardiac arrest.

Для цитирования: Cardim Danilo, Griesdale Donald E, Ainslie Philip N, Robba Chiara, Calviello Leanne, Czosnyka Marek, Smielewski Peter, Sekhon Mypinder S. A comparison of non-invasive versus invasive measures of intracranial pressure in hypoxic ischaemic brain injury after cardiac arrest.. Resuscitation. 2019. DOI: 10.1016/j.resuscitation.2019.01.002.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.