DOI 10.1097/ijg.0000000000000324

Медицинская библиотека · Journal of glaucoma · 2016

Initial Experience With the New Ahmed Glaucoma Valve Model M4: Short-term Results.

Cvintal Victor, Moster Marlene R, Shyu Andrew P, McDermott Katie, Ekici Feyzahan, Pro Michael J, Waisbourd Michael

СтатьяПеревод в очереди

Аннотация

PURPOSE: To evaluate the clinical outcomes of the new Ahmed glaucoma valve (AGV) model M4. The device consists of a porous polyethylene shell designed for improved tissue integration and reduced encapsulation of the plate for better intraocular pressure (IOP) control.

METHODS

Medical records of patients with an AGV M4 implantation between December 1, 2012 and December 31, 2013 were reviewed. The main outcome measure was surgical failure, defined as either (1) IOP<5 mm Hg or >21 mm Hg and/or <20% reduction of IOP at last follow-up visit, (2) a reoperation for glaucoma, and/or (3) loss of light perception.

RESULTS

Seventy-five eyes of 73 patients were included. Postoperative IOP at all follow-up visits significantly decreased from a baseline IOP of 31.2 mm Hg (P<0.01). However, IOP increased significantly at 3 months (20.4 mm Hg), 6 months (19.3 mm Hg), and 12 months (20.3 mm Hg) compared with 1 month (13.8 mm Hg) postoperatively (P<0.05). At 6 months and 1 year, the cumulative probability of failure was 32% and 72%, respectively.

CONCLUSIONS

The AGV M4 effectively reduced IOP in the first postoperative month, but IOP steadily increased thereafter. Consequently, failure rates were high after 1 year of follow-up.

Для цитирования: Cvintal Victor, Moster Marlene R, Shyu Andrew P, McDermott Katie, Ekici Feyzahan, Pro Michael J, Waisbourd Michael. Initial Experience With the New Ahmed Glaucoma Valve Model M4: Short-term Results.. Journal of glaucoma. 2016. DOI: 10.1097/ijg.0000000000000324.