Validation of the automated oscillometric home blood pressure monitor AViTA BPM82 in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2:2018).

СтатьяВалидационное исследованиеНазвание и аннотация: Перевод в очереди
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Menti Ariadni, Lempidaki Petroula, Mariglis Dimitrios, Antonogiannaki Eirini Chrysovalanti, Bakalakou Konstantina-Rozalia, Komnianou Aikaterini, Kollias Anastasios, Stergiou George S

Аннотация

OBJECTIVE: To evaluate the accuracy of the automated upper-arm cuff oscillometric home blood pressure (BP) monitor AViTA BPM82 in a general population according to the Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) Universal Standard (ISO 81060-2 : 2018) and its amendments (1 : 2020 and 2 : 2024).

METHODS: Participants were recruited to fulfill the age, sex, BP, and arm distribution criteria of the AAMI/ESH/ISO Universal Standard (ISO 81060-2 : 2018) and its amendments (1 : 2020 and 2 : 2024) in a general population using the same arm sequential measurement method. A single 360° wide-range hard cuff of the test device was used for arm circumference 22-42 cm.

RESULTS: One-hundred-twelve individuals were recruited and 85 were analyzed [mean age 58.1 ± 15.2 (SD) years, 34 men, arm circumference 32 ± 5.5 cm, range: 22-42 cm]. For validation of Criterion 1, the mean difference ± SD between the test device and reference BP readings ( N  255) was -1.6 ± 7.4/-1.5 ± 5.6 mmHg (systolic/diastolic; threshold: ≤5 ± 8 mmHg). For Criterion 2, the SD of the averaged BP differences between the test device and reference BP per individual ( N  85) was 6.20/5.11 mmHg (systolic/diastolic; threshold: ≤6.76/6.78 mmHg).

CONCLUSION: The automated oscillometric home BP monitor AViTA BPM82 fulfilled all the requirements of the AAMI/ESH/ISO Universal Standard (ISO 81060-2 : 2018) and its amendments (1 : 2020 and 2 : 2024) in a general population and can be recommended for home use.