A New Tool to Assess Patient-Ventilator Synchrony in Preterm Infants Receiving Non-Invasive Ventilation: A Randomized Crossover Pilot Study.

Cresi Francesco, Maggiora Elena, Rubino Carlotta, Ferroglio Mattia, Ruzzante Elena, Piga Enrico, Giraudo Isaac, Limone Marco, Terrin Gianluca, Coscia Alessandra

СтатьяРандомизированное исследованиеПеревод в очереди

Аннотация

INTRODUCTION

Nasal synchronized intermittent positive pressure ventilation (nSIPPV) is an effective non-invasive ventilation technique, especially for preterm infants. Patient-ventilator synchrony is essential for providing effective respiratory support; however, no automated system is currently available for monitoring this parameter. A new tool for automatic assessment of patient-ventilator synchrony, the SyncNIV system, was developed and applied in this pilot study to evaluate differences between nSIPPV and non-synchronized nasal intermittent positive pressure ventilation (nIPPV) in preterm infants with respiratory distress.

METHODS

This study involved designing a custom algorithm for signal analysis. Data were collected through a polygraph that could simultaneously gather respiratory data from the patients and the ventilator. Patient-ventilator synchrony was evaluated by applying the SyncNIV system in a randomized crossover study designed to compare nSIPPV and nIPPV. The primary outcome was the mean instant Synchrony Index (i-SI), defined as the portion of the inspiration effort sustained by ventilator inflation, expressed as a percentage.

RESULTS

Fourteen infants with a median (IQR) gestational age of 28.6 (25.6-30.3) were enrolled. We analyzed 43,304 ventilator inflations and 50,221 patient breaths. The i-SI was 54.69% (44.49-60.09) in nSIPPV and 39.54% (33.40-48.75) in nIPPV, p < 0.05.

CONCLUSION

The SyncNIV system confirmed better i-SI during nSIPPV than during nIPPV, demonstrating its effectiveness in assessing the differences between these two modes of non-invasive ventilation in preterm infants. The SyncNIV system could be a useful tool for optimizing the ventilation parameters and improving the effectiveness and comfort of respiratory support systems.

Для цитирования: Cresi Francesco, Maggiora Elena, Rubino Carlotta, Ferroglio Mattia, Ruzzante Elena, Piga Enrico, Giraudo Isaac, Limone Marco, Terrin Gianluca, Coscia Alessandra. A New Tool to Assess Patient-Ventilator Synchrony in Preterm Infants Receiving Non-Invasive Ventilation: A Randomized Crossover Pilot Study.. Neonatology. 2025. DOI: 10.1159/000543413.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.