Beware of the B(e)all valve: mistaken valve identity, 30-year survival, and valve replacement.
Аннотация
Effective management of patients after the implantation of mechanical cardiac valves includes correct recognition of each valve and its related complications. Herein, we present the case of a patient who had undergone implantation of a floating-disc Beall-Surgitool mitral valve in 1976 and developed multiple valve-related complications. Over 30 years and in multiple medical centers, the device was mistakenly assumed to be a "ball" valve. The correct identification of the prosthesis led to the recognition of valvular failure, and the patient underwent its replacement with an On-X bileaflet carbon valve. Pathologic and microscopic examination of the explanted Beall valve showed massive pannus formation that extended over the sewing cuff on the atrial and ventricular side, preventing complete disc closure; disrupted fabric coating of the sewing ring, with exposure of the underlying metal; and a marked inflammatory reaction. We report one of the longest intervals on record between the implantation and replacement of a Beall-Surgitool valve.