MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1109/tnsre.2026.3721183

Медицинская библиотека · IEEE transactions on neural systems and rehabilitation engineering : a publication of the IEEE Engineering in Medicine and Biology Society · 2026

Phase-Synchronized TECS Integrating TI-tACS and rTMS: Modeling, Hardware Implementation, and In Vivo Neurophysiological Validation.

Wang Tingyu, Guo Miaomiao, Cui Chao, Wei Yujia, Zhao Xiuzhi, Zheng Weiran, Wang Alan, Xu Guizhi

СтатьяПеревод в очереди

Аннотация

OBJECTIVE

Conventional noninvasive neuromodulation is constrained by trade-offs among field strength, focality, and penetration. We propose a phase-synchronized transcranial electromagnetic combined stimulation (TECS) approach that concurrently combines temporal interference transcranial alternating current stimulation (TI-tACS) with repetitive transcranial magnetic stimulation (rTMS) to improve the local intensity-focality trade-off in the motor cortex. To our knowledge, this is the first study to investigate a phase-synchronized concurrent TI-tACS + rTMS paradigm.

METHODS

Finite-element simulations quantified the peak magnitude-based stimulation metric in the target M1 region, together with the equivalent focal area and effective penetration depth under a half-maximum criterion. A phase-synchronized TECS system was implemented and evaluated in rats using rotarod performance, c-Fos immunofluorescence, and histology, with comparisons to TI-tACS alone, rTMS alone, and a shamlike control condition.

RESULTS

TECS produced the highest peak target-region electric-field value (178.0 V/m) while maintaining a much smaller focal area than rTMS alone (8.3 vs. 76.0 mm2) and greater penetration than TI-tACS alone (2.1 vs. 1.2 mm). In vivo, TECS produced the longest rotarod latency on Day ${7}({288.5} \pm {16.4} {~\text {s}}; \text {n}={6}$ rats/group; daily values averaged over three trials/rat), exceeding the control, TItACS, and rTMS groups by 75.3%, 15.3%, and 19.4%, respectively. TECS also induced the highest c-Fos activation, with corresponding increases of 50.1%, 9.4%, and 12.8%, respectively. Histology revealed no overt abnormalities across conditions.

CONCLUSION

Phase-synchronized TECS enhances focal electric-field intensity and neuronal activation, translating to improved motor performance.

SIGNIFICANCE: The proposed TECS platform provides a promising neuroengineering approach for targeted noninvasive brain stimulation, supported by simulationdefined field metrics and in vivo feasibility and safety validation.

Для цитирования: Wang Tingyu, Guo Miaomiao, Cui Chao, Wei Yujia, Zhao Xiuzhi, Zheng Weiran, Wang Alan, Xu Guizhi. Phase-Synchronized TECS Integrating TI-tACS and rTMS: Modeling, Hardware Implementation, and In Vivo Neurophysiological Validation.. IEEE transactions on neural systems and rehabilitation engineering : a publication of the IEEE Engineering in Medicine and Biology Society. 2026. DOI: 10.1109/tnsre.2026.3721183.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.