MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1109/embc.2013.6609854

Медицинская библиотека · Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference · 2013

Esophageal SpO2 measurements from a pediatric burns-patient: a case study.

Kyriacou P A, May J M, Petros A J

Клинический случайСтатьяПоддержка исследования: вне правительства СШАПеревод в очереди

Аннотация

Pulse oximetry is being used in everyday clinical practice in anesthesia utilizing peripheral saturation sensors. However, it may be unreliable in certain clinical situations such as peripheral hypoperfusion. Similar situations occur in burns patients and more importantly burns to extremities which limit the sites available for measurement of peripheral oxygen saturation (SpO2). To overcome these limitations, the esophagus has been investigated as an alternative measurement site, as perfusion may be preferentially preserved centrally. A miniaturized reflectance esophageal saturation (SpO2 probe has been constructed utilizing infrared and red photodiodes and a photodetector. Our case study was aimed at evaluating the reliability of esophageal pulse oximetry in a major burns infant. Measurable photoplethysmographic (PPG) traces and SpO2 values were obtained in the neonatal esophagus. It was found that the esophageal pulse oximeter results were in good agreement with oxygen saturation measurements obtained by a commercial ear lobe pulse oximeter. This study suggests that the esophagus can be used as an alternative site for monitoring arterial blood oxygen saturation by pulse oximetry in burned infants.

Для цитирования: Kyriacou P A, May J M, Petros A J. Esophageal SpO2 measurements from a pediatric burns-patient: a case study.. Annual International Conference of the IEEE Engineering in Medicine and Biology Society. IEEE Engineering in Medicine and Biology Society. Annual International Conference. 2013. DOI: 10.1109/embc.2013.6609854.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.