Straight video blades are advantageous than curved blades in simulated pediatric difficult intubation.
Аннотация
BACKGROUND
It is still controversial which laryngoscope may be a better option in unanticipated difficult airway in pediatric patients. The aim of the present study was to compare two direct and two video-assisted laryngoscope devices for the management of difficult pediatric airway.
METHODS
Forty-five anesthesiology residents and nurse anesthetists participated in the study. Macintosh, Miller, Storz Miller, and McGrath Mac curved laryngoscopes were used for tracheal intubation of 3-6-month Airsim Pierre Robin manikin. We compared the duration of successful intubation, number of attempts, glottic view grades, severity of dental trauma, the use of optimization maneuvers, and the difficulty of use of the devices with straight and curved laryngoscope blades.
RESULTS
Successful intubation duration was significantly lower in Storz Miller device, and the number of intubation attempts was significantly higher in the Macintosh laryngoscope (P < 0.01). According to the Cormack and Lehane classification, Grades 1 and 2 glottic view was 20% for Macintosh and 40% for Miller laryngoscope, while it was 100% for Storz Miller and 80% for McGrath (P < 0.001). Difficulty VAS scores of Storz Miller device were significantly lower than the scores of Macintosh, Miller, and McGrath laryngoscopes (15.7 ± 14.89, 34.7 ± 26.44, 31.5 ± 26.74, 33.4 ± 26.67 mm, respectively; P < 0.01). The severity of dental trauma was significantly lower in Storz Miller compared with Macintosh, Miller, and McGrath laryngoscopes (0.96 ± 1.04, 1.67 ± 1.15, 1.38 ± 1.05, 1.42 ± 1.27, respectively; P < 0.01).
CONCLUSION
Storz Miller laryngoscope was found to have advantages over the other laryngoscopes in regard to glottic view, duration of successful intubation, number of attempts, dental trauma severity, need for additional maneuvers, and ease of use.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.